![Fresenius Medical Care multiFiltrate Скачать руководство пользователя страница 172](http://html.mh-extra.com/html/fresenius-medical-care/multifiltrate/multifiltrate_instructions-for-use-manual_2331671172.webp)
Chapter 5: Alarm processing
5-6
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
5.6
Alarm system response
When starting treatment or resuming treatment after an alarm
When starting treatment or resuming treatment after an alarm, you can
delay the following alarms or widen the alarm limits for a certain period
of time by pressing the
START/RESET
key.
Overriding an alarm (temporarily deactivating an alarm system)
If a pending blood leak alarm is confirmed with the
START/RESET
key,
this alarm is overridden (deactivated) for a certain period of time.
Suppressing the audible alarm (mute)
The
Mute
key allows the operator to suppress (deactivate) the audible
tone of a signaled alarm for a certain period of time.
If a new alarm occurs during this time, the audible alarm tone of the new
alarm is signaled regardless.
Alarm
Response
Arterial pressure
Alarm limits widened to maximum values for
the duration of the pump start-up
(approx. 10 seconds)
Venous pressure
Alarm limits widened to maximum values for
the duration of the pump start-up
(approx. 10 seconds)
TMP
Alarm limits widened to maximum values for
the duration of the pump start-up
(approx. 10 seconds)
pF pressure
Alarm limits widened to maximum values for
the duration of the pump start-up
(approx. 10 seconds)
Blood leak
Start delayed for approx. 8 seconds
Warning
While overriding the alarm system, the operator is responsible for the
patient’s safety.
Alarm
Override time
Blood leak
2 minutes
Содержание multiFiltrate
Страница 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Страница 2: ......
Страница 11: ...Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 xi 15 2 Instructions on the use of free software 15 2 ...
Страница 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 166: ...Chapter 4 Operation 4 124 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 Interrupting the treatment ...
Страница 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 234: ...Chapter 11 Technical Safety Checks maintenance procedures 11 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Страница 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...