Chapter 4: Operation
60
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
4.8 EMERGENCY MODE operating state
4.8.1 General information
Operating modes–overview
Emergency mode screen
Warning
Unpredictable device response
Some essential program functions are inactive during emergency
mode.
EMERGENCY MODE
is only designed for short-term use (e.g., to
complete the dialysis treatment in progress; maximum 120 hours).
The issue which caused the device to switch to emergency mode
must be eliminated immediately to resume normal operation.
Warning
Emergency mode after a disinfection
The
EMERGENCY MODE
may not be started if there is still residual
disinfectant in the system after a disinfection.
During
EMERGENCY MODE
the electronic
control is bypassed.
EMERGENCY MODE
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...