Chapter 6: Cleaning, disinfection, preservation
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
135
6.3 Disinfection
6.3.1
General notes
Operating principle
The disinfection procedure is program-controlled.
Reason for a disinfection
– If a water supply as specified by the applicable regulations can no
longer be ensured:
– After
repairs
to the dialysis water circuit.
– If the system has been
idle
for more than
72 hours
. In the event of
longer periods without use, preservation of the device is
recommended.
– ISO 23500-1 (or ISO 23500 respectively) "Guidance for the
preparation and quality management of fluids for haemodialysis and
related therapies" recommends a regular (e.g., monthly)
preventive
disinfection
to avoid significant biofilm formation (biofouling).
Recommended
disinfectant
Minncare
®
6.3.2
Disinfecting the system
Warning
Operator instructions
The device may only be cleaned, disinfected, and preserved by persons
who have been instructed on the proper handling of the device during
such procedures.
The operator must observe and follow the general safety
precautions.
The system may only be disinfected after consultation with the
manufacturer of the system or by persons authorized by the
manufacturer.
Note
If the microbiological tests indicate a persistent microbial count in the
dialysis water, shorten the disinfection interval.
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...