5.7 Error category 04–Start test and test routines
An error code marked with „*“ denotes a self-confirming message.
Error code
Visual
indica-
tor
Messages
Cause
Action required
F–04–01–01
Signal:
red
FAILURE:
T1 Test
Measuring
transducer function
not ensured
– Reference measurement of
test voltage (8.0 V
DC
) failed
– Measuring transducer B4
defective
– Serial connection line
COM1 defective
– Digital output terminal A13
defective
– Line connection between
measuring transducer B4
and analog output terminal
A13 defective
Contact the service
department
F–04–01–02
Signal:
red
FAILURE:
T1 Test
Temperature
measurement,
function not ensured
– Deviation between T-F and
T-P more than 5 °C
– Temperature sensor T-F
and T-P defective
– Deviation between T-P and
T-Ps more than 5 °C
(for
AquaA2
only)
– Temperature sensor T-Ps
defective (for
AquaA2
only)
Contact the service
department
F–04–01–04
Signal:
red
FAILURE:
T1 Test
Booster pump 1;
Function not
ensured
– The booster pump 1 fails to
build up concentrate
pressure.
– P-C sensor defective
– The motor protection switch
F2 has tripped.
– Measuring transducer B4
defective
Check the pump
Contact the service
department
F–04–01–06
Signal:
red
FAILURE:
Permeate
conductivity cell
defective
– The line connection to the
dialysis water conductivity
sensor is defective or
interrupted.
– Conductivity sensor CD-P
defective
– Measuring transducer B4
defective
Contact the service
department
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...