5.6 Error category 03–start condition not fulfilled
An error code marked with „*“ denotes a self-confirming message.
Error code
Visual
indica-
tor
Messages
Cause
Action required
W–03–01–01*
Signal:
yellow
WARNING:
Rinse start, tank
cannot be filled
– Failure to reach
NIV3
.
– Water inlet pressure too low
Check water supply
Contact the service
department
W–03–01–02*
Signal:
yellow
WARNING:
Rinse start,
pressure cannot be
built up
– The concentrate pressure
failed to rise above the
specified limit.
– Pressure sensor P-C
defective
– Measuring transducer B4
defective
Check the pumps
Contact the service
department
W–03–01–03*
Signal:
yellow
WARNING:
Rinse start,
Operating point
(pressure) not
reached
– Booster pump P1 defective
– The motor protection switch
F1 has tripped.
– The concentrate pressure
failed to rise above the
specified limit.
– Pressure sensor P-C
defective
– Measuring transducer B4
defective
Check the pumps
Contact the service
department
W–03–01–04*
Signal:
yellow
WARNING:
Rinse start, no
circulation flow
– Flow control switch of pump
P3 defective
– Circulation pump P3
defective
– The motor protection switch
F3 has tripped.
The circulation pump P3
fails to deliver
Check the pumps
Contact the service
department
W–03–01–05*
Signal:
yellow
WARNING:
Rinse start,
permeate cond. too
high
– The dialysis water
conductivity CD-P failed to
drop below the specified
alarm limit.
– Conductivity sensor CDT-P
defective
– Measuring transducer B4
defective
Contact the service
department
W–03–01–06*
Signal:
yellow
WARNING:
Start, tank cannot
be filled
– Failure to reach
NIV3
.
– Water inlet pressure too low
Check water supply
Contact the service
department
Содержание AquaA
Страница 2: ......
Страница 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 138: ...Chapter 6 Cleaning disinfection preservation 138 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 156: ...Chapter 11 Technical Safety Checks and maintenance 156 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Страница 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Страница 228: ...Chapter 15 Appendix 228 Fresenius Medical Care AquaA IFU EN US 06B 2021 Participant Name Date signature ...