Chapter 4: Installation
4-12
Fresenius Medical Care 5008 SM 1/08.04
8
Check – dialysate pressure
In the service mode, select CALIBRATE.
Basic requirements:
The hemodialysis system
must be closed.
The reference measuring instrument must be placed at the bottommost position of the IV pole.
Dialyzer couplings must be connected to the reference instrument.
Flow on until dialysate lines and reference measuring instrument are free from air. Then flow off.
8.1
Zero point S03/S07
Reference measuring instrument: 0 bar
Open the vent valve (UMED).
Using a syringe (filled with fluid) set a value of 0 bar, via the vent valve.
Check S03.
Desired value: +16 mbar to +76 mbar
Check S07.
Desired value: +16 mbar to +76 mbar
8.2
Slope S03/S07 (+)
Reference measuring instrument: +533 mbar (
±
26 mbar)
Using a syringe (filled with fluid) set a value of +533 bar, via the vent valve.
Check S03.
Desired value: S03 = Display of reference measuring inst (+16 mbar to +76 mbar)
Check S07.
Desired value: S07 = Display of reference measuring inst (+16 mbar to +76 mbar)
8.3
Slope S03/S07 (–)
Reference measuring instrument: –533 mbar (
±
26 mbar)
Using a syringe (filled with fluid) set a value of –533 bar, via the vent valve.
Check S03.
Desired value: S03 = Display of reference measuring inst (+16 mbar to +76 mbar)
Check S07.
Desired value: S07 = Display of reference measuring inst (+16 mbar to +76 mbar)
No.
Description
Содержание 5008
Страница 2: ......
Страница 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 100: ...Chapter 8 Tools Service Equipment 8 4 Fresenius Medical Care 5008 SM 1 08 04 Toolcase 5008 Part number M35 463 1 ...
Страница 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Страница 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...