Chapter 3: Specifications
Fresenius Medical Care 5008 SM 1/08.04
3-5
Cleaning programs
Rinse clear/rinse/mandatory rinse:
Time adjustable in the TECHNICIAN's SETUP,
Temperature: approx. 37 °C,
Flow: 600, 800 ml/min (adjustable in the SETUP)
Cold disinfection/degreasing, cold disinfection:
Time adjustable in the TECHNICIAN's SETUP,
Temperature: approx. 37 °C,
Flow: max. 900 ml/min
Heat disinfection:
Time adjustable in the TECHNICIAN's SETUP,
Flow: max. 900 ml/min
In all programs:
Blood pump stops, arterial and venous line occlusion clamp closed.
Progress of the program (time-counting) is interrupted in the event of a
flow alarm.
The cleaning programs can be aborted.
The chemical disinfection program is followed by a mandatory rinse.
Flush
Rinsing of the water supply area
O
Dialysate circuit and safety systems
Blood leak detector
Threshold of response
≤
0.5 ml blood loss per minute into the dialysate
at a hematocrit of 0.25.
(flow rate 100 ml/min to 1000 ml/min)
Transmembrane pressure
Display range: –100 to 400 mmHg
Resolution: 5 mmHg
Definition:
Ultrafiltration
Selectable UF rate: 0 ml/h to 4000 ml/h (in 10 ml increments)
Maximum rate internally adjustable to 1, 2, 3, or 4 l/h.
Pump volume accuracy: ±1 % (at P
di
> –500 mbar)
The UF rate/effective blood flow ratio is being monitored during the
treatment. If an incongruity occurs a warning will be displayed after
approx. 10 seconds.
Pressure holding test
Event-controlled
TMP
= P
bo
– (P
di
+ P
do
) / 2 + Offset
TMP
= Transmembrane pressure
P
bo
= Blood pressure on the outlet side of the dialyzer
P
di
= Dialysate pressure on the inlet side of the dialyzer
P
do
= Dialysate pressure on the outlet side of the dialyzer
Offset = Flow-dependent pressure fluctuations
Содержание 5008
Страница 2: ......
Страница 6: ...Kapitel 1 Index 1 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 10: ...Chapter 2 Important Information 2 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 22: ...Chapter 3 Specifications 3 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 52: ...Chapter 5 Setup 5 12 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 60: ...Chapter 5 Setup 5 20 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 66: ...Chapter 6 TSC TMC Maintenance 6 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 74: ...Chapter 6 TSC TMC Maintenance 6 10 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 90: ...Chapter 6 TSC TMC Maintenance 6 26 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 96: ...Chapter 7 Error Messages 7 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 100: ...Chapter 8 Tools Service Equipment 8 4 Fresenius Medical Care 5008 SM 1 08 04 Toolcase 5008 Part number M35 463 1 ...
Страница 107: ...Chapter 10 Repair Fresenius Medical Care 5008 SM 1 08 04 10 1 10 Repair no contents yet ...
Страница 108: ...Chapter 10 Repair 10 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 110: ...Chapter 11 Functional Description 11 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 128: ...Chapter 12 Service Program Option 12 2 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 130: ...Chapter 12 Service Program Option 12 4 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 132: ...Chapter 12 Service Program Option 12 6 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 150: ...Chapter 12 Service Program Option 12 24 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 164: ...Chapter 12 Service Program Option 12 38 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 166: ...Chapter 12 Service Program Option 12 40 Fresenius Medical Care 5008 SM 1 08 04 ...
Страница 168: ...Chapter 12 Service Program Option 12 42 Fresenius Medical Care 5008 SM 1 08 04 ...