M A I N T E N A N C E
C o n t i n u e d
P/N 460018 Rev. D
25
7.
Checking bleach residuals during DRAIN Operation from the drain hose.
•
Have chlorine test strip ready for use.
•
Using STEP button step to the DRAIN Operation.
•
Press the RINSE CYCLE side START button.
•
PAUSE Indicator Light will turn off and the Drain Valve will OPEN. Allow water to drain for 30 seconds.
•
Place test strip under the flow for 30 seconds or the recommended time on the test strip
manufacture’s instruction. Press PAUSE button once you are complete with this step.
•
Ensure that you check the results of the test strip immediately.
8.
If residual bleach levels are higher than ANSI/AAMI Standard limit of <0.1 ppm (RD61: 2006), in any of the
two testing regions initiate another complete RINSE CYCLE. After the RINSE CYCLE is complete, start from
section 8.3, #5, to check for residual bleach. Continue the RINSE CYCLE and test procedure until bleach
residuals are within ANSI/AAMI Standard limit of <0.1 ppm (RD61:2006) at the end of the Transfer and
Drain Hose.
9.
Once you have attained an acceptable bleach residual reading, ensure the Mixing Tank is empty, power to
the Dry Acid Dissolution Unit has been turned off and CLOSE the Main Transfer Ball Valve near the filter
housing. Then, connect the Transfer Hose back onto the Dry Acid Dissolution Unit Transfer Hose Holder
(See Diagram 2: Dry Acid Dissolution Unit, page 8).
10.
Remove Filter Housing and drain all residual water from the Housing. Install new Filter and tighten Filter
Housing into place (Figure 21).
Figure 21
11.
Immediately after RINSE CYCLE, make a batch of Dry Acid Product using Section 7, page 14. Leaving the
Dry Acid Dissolution with only treated water or wetted with only treated water leaves the Unit
susceptible to bacterial growth.
8.4
F
ILTER
M
AINTENANCE
The 1-micron filter should be changed under the following conditions:
1. The Filter should be changed after mixing 6 batches.
2. When the Dry Acid Dissolution Unit requires disinfection.
CAUTION!
D
O NOT ALLOW THE
U
NIT TO REMAIN FULL OF WATER WITHOUT THE ADDITION OF
F
RESENIUS
M
EDICAL
C
ARE
D
RY
A
CID
P
RODUCT
.
B
ACTERIAL GROWTH MAY OCCUR
.
A. Remove filter housing
B. Insert filter (P/N G84-
202-12)
C. Install filter housing onto
the filter holder.
D. Tighten the filter housing
Содержание 460018
Страница 1: ... DRY ACID DISSOLUTION UNIT 132 GALLONS 500 LITERS OPERATOR S MANUAL P N 460018 REV D ...
Страница 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 37: ...P N 460018 Rev D 31 PAGE INTENTIONALLY LEFT BLANK ...
Страница 39: ...P N 460018 Rev D 33 PAGE INTENTIONALLY LEFT BLANK ...
Страница 41: ...P N 460018 Rev D 35 PAGE INTENTIONALLY LEFT BLANK ...
Страница 43: ...P N 460018 Rev D 37 PAGE INTENTIONALLY LEFT BLANK ...
Страница 45: ...P N 460018 Rev D 39 PAGE INTENTIONALLY LEFT BLANK ...
Страница 50: ...44 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 51: ...P N 460018 Rev D 45 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Страница 52: ...46 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 53: ...P N 460018 Rev D 47 ...
Страница 54: ...48 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 55: ...P N 460018 Rev D 49 ...
Страница 56: ...50 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 57: ...P N 460018 Rev D 51 ...
Страница 58: ...52 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 59: ...P N 460018 Rev D 53 ...
Страница 60: ...54 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 61: ...P N 460018 Rev D 55 ...
Страница 62: ...56 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 63: ...P N 460018 Rev D 57 ...
Страница 64: ...58 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 65: ...P N 460018 Rev D 59 ...
Страница 66: ...60 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 68: ...62 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 70: ...64 P N 460018 Rev D PAGE INTENTIONALLY LEFT BLANK ...
Страница 75: ...P N 460018 Rev D 69 PAGE INTENTIONALLY LEFT BLANK ...