Fresenius Medical Care 4008 HDF Скачать руководство пользователя страница 1

Care

Fresenius Medical

4008 HDF

Technical Manual

Содержание 4008 HDF

Страница 1: ...Care Fresenius Medical 4008 HDF Technical Manual ...

Страница 2: ...4008 HDF Technical Manual ...

Страница 3: ... of EN 60601 1 Assembly extension adjustment modification or repair may only be carried out by the manufacturer or persons authorized by him Any inquiries should be addressed to Manufacturer Fresenius Medical Care Deutschland GmbH Plant Schweinfurt Hafenstraße 9 D 97424 Schweinfurt Germany Tel 09721 678 0 Fax 09721 678 200 Head office Fresenius Medical Care Deutschland GmbH Borkenberg 14 D 61440 O...

Страница 4: ...0 2 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 5: ...l are available under part no 674 406 1 who have the necessary background experience in mechanics electrical and medical engineering who have been authorized by the manufacturer to perform maintenance and repair work who have access to the necessary auxiliary and measuring equipment Restrictions The study of this manual does not represent an alternative to the training courses offered by the manuf...

Страница 6: ...tion number 75 with a resistance of 1 5 Ohm The decimal point is replaced by a unit symbol to reduce the possibility of errors Resistors Capacitors R1 0 1 Ω µ 1 0 1 µF 1R5 1 5 Ω 1µ5 1 5 µF 1K5 1 5 kΩ 1000µ 1000 µF Note When repairing or exchanging replacement parts make sure to take the applica ble ESD precautions e g EN 100 015 1 During repair troubleshooting in the hydraulic unit protect the com...

Страница 7: ...ns 3 7 3 3 Calibrating the 4008 HDF scale 3 8 3 4 Calibrating the substituate sensor 3 13 3 5 Calibrating the HDF blood pump 3 15 3 6 Repair instructions 3 17 4 Circuit descriptions and circuit diagrams 4 4 1 LP 625 display board 4 3 4 2 LP 754 control board HDF 4 9 4 3 LP 760 HDF motor control 4 19 4 4 LP 761 HDF drive keyboard 4 27 4 5 LP 762 scale comm 4 33 4 6 LP 763 SSE serial interface exten...

Страница 8: ...0 6 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 9: ... 1 3 1 1 1 Components 1 3 1 1 2 Description 1 3 1 1 3 Component tests 1 3 Fig Block diagram 1 4 Fig Flow diagram 1 5 Fig Flow diagram 1 6 1 2 Description of extended T1 test error messages in T1 test 1 8 1 2 1 Test UF Function 1 8 1 3 Error messages during treatment and the HDF test 1 10 1 4 Error messages substituate pump 1 13 ...

Страница 10: ...1 2 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 11: ... A substituate sensor mounted on the substituate lift recognises whether Line set is fitted without substituate or Line set is fitted with substituate The substituate pump delivers the substituate solution to the venous bubble catcher The filtrate from the dialysis fluid circuit is removed by the UF2 pump 1 1 3 Component tests The scales are checked by the operator by means of a plausibility test ...

Страница 12: ...Fresenius Medical Care 4008 HDF 3 11 97 TM Scales CPU 1 UF2 pump Substituate pump TMP CPU 2 W Serial communication Serial communication Serial communication Operating system Safety system Fig Block diagram ...

Страница 13: ...P 631 CPU 1 LP 763 SSE Serial interface extender LP 633 Input board LP 634 Output board serial interface X632 B24 X632 C27 X634 R C11 X634 R A24 serial interface serial interface and taring line X632 A7 X632 C7 X632 A29 X634 L A C24 X634 L A C23 X633 L C14 X633 L C23 X633 L B6 Control UF2 pump Control UF pump Acknowledgement dialysate pressure Control UF pump Control UF2 pump Acknowledgement UF pu...

Страница 14: ...1 6 Fresenius Medical Care 4008 HDF 3 11 97 TM Fig Flow diagram ...

Страница 15: ...Hydraulic measurement points A Reduced water inlet pressure B Balancing chamber loading pressure C Flow pump pressure D Degassing pump pressure 77 Heat exchanger 78 Relief valve 84 Disinfection valve 85 Disinfectant conncector 86 Recirculation valve 87 Discharge valve 88 Multifunction block 88a Degassing chamber 88b Secondary air separator 88c Primary air separator 89 Degassing orifice 90a Rinse c...

Страница 16: ...chnical Manual for the hemodialysis machine If the HDF option is installed and activated the following error codes can also appear 1 2 1 Test UF Function Test description CPU 1 starts up the UF2 pump at a defined rate CPU 2 controls the hydraulic and electrical function of the UF2 pump CPU 2 blocks the control line of the UF2 pump and checks for standstill UF2 counter check Figure ...

Страница 17: ...mum rate of 5000 ml h is not possible Monoflop on LP 634 defect IC 42 R65 C45 F14 UF Function UF2 pump not active during the test 4 s Acknowledgement UF_P2 X637 B26 X632 C7 no LOW pulse Control line UF_P2 X634L A C24 X637 B26 no LOW pulse F15 UF Function The UF2 pump cannot be stopped by the CPU2 Control line UF_P_EN X632 C28 X634R A22 no 5 V Reset input to IC 42 pin 13 on LP 634 defect F16 UF Fun...

Страница 18: ...d Air will still be detected after 2 minutes Since fluid has been recog nized in the secondary air sep arator the substituate pump starts at a rate of 10 ml min until a change in weight follows on the scales i e a jump in grammes There was no dif ference in weight established after one minute If the expected change in weight i e a jump in grammes is higher or lower than 1 g the respective UF pump ...

Страница 19: ... 10 seconds Difference between desired or actual delivery rate of the UF pump greater than 10 UF pump stops longer than the maximum period time Despite switched off ultrafiltra tion the change in delivery rate of the UF pump is greater than 10 ml Interval between two strokes of the UF2 pump less than 220ms Check substituate hose clamped off not connected with the venous bubble catcher Check UF pum...

Страница 20: ... 180 ms Pulse time of a stroke of the UF2 pump more than 500 ms Reaction time of the UF2 pump more than 10 seconds Difference in desired or actual delivery rate of UF2 pump greater than 10 UF2 pump stops longer than the maximum period time Despite switched off 4008HDF change in delivery rate of the UF2 pump greater than 10 ml Failure of UF pump Failure of UF2 pump Volume difference between UF pump...

Страница 21: ...ion E 05 SN stroke volume outside the permitted range E 06 The SN pressure thresholds outside the value range of the AD converter E 07 Not defined E 08 Failure in the AD conversion E 09 Not defined E 10 Not defined E 11 Not defined E 12 Failure speed monitoring hall sensor E 13 Failure monitoring system current sensing resistors E 14 Failure monitoring system current sensing resistors E 15 Failure...

Страница 22: ...1 14 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 23: ...ety checks TSC These inspections must be carried out every 12 months The technical safety inspections stipulated for the hemodialysis machine must be carried out in addition to these technical saftey checks The technical safety checks are to be recorded in the equipment log A technical safety checks report is to be found at the page 2 3 ...

Страница 24: ...2 2 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 25: ...s 1 UF2 pump TSC 1 1 Delivery volume 1 ml stroke 60 strokes 60 ml 1 2 VDE inspections values according to EN 60601 1 TSC 2 1 Protective ground resistance max 0 3 Ω TSC 2 2 Summarized leakage current Must fulfill both conditions 1 No more than 1 5 times the summari zed value first measured Summarized value first measured Refer to the machine card enclosed with the machine 2 No more than 1 mA The te...

Страница 26: ...2 4 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 27: ...mation on the calibration instructions 3 3 3 1 Calibrating the UF2 pump 3 5 3 2 VDE inspections 3 7 3 3 Calibrating the 4008 HDF scales 3 8 3 4 Calibrating the substituate sensor 3 13 3 5 Calibrating the HDF blood pump 3 15 3 5 1 Pressure transducer 3 15 3 5 2 Checking SN switchover points 3 16 3 6 Repair instructions 3 17 ...

Страница 28: ...3 2 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 29: ...rt Stop Rate ml min Ø mm Start Stop Rate ml min Ø mm HDF Reset Vol l Pven Start Stop Rate ml h Bolus ml Bolus Rate h min 40 0 100 200 300 100 200 280 0 10 20 30 10 20 30 0 100 200 0 10 20 30 300 400 500 40 50 60 0 100 200 0 10 20 30 300 400 500 40 50 60 13 13 5 14 14 5 15 15 5 300 500 700 900 35 37 39 Fresenius Medical Care Fresenius Medical Care LP 763 LP 762 LP 761 LP 625 LP 754 LP 760 3 0 Gener...

Страница 30: ...ius Medical Care 4008 HDF 3 11 97 TM 1 2 3 4 5 6 7 8 ON OFF LP 632 CPU 2 SW1 Fig DIP switch 4008 HDF SW 1 DIP switch 8 ON 4008 HDF test selectable required not required OFF 4008 HDF test carried out automatically ...

Страница 31: ...librating the UF2 pump Switch off hemodialysis machine Set service switch to ON top Switch on hemodialysis machine Confirm Select CALIBRATION Confirm Select ADJ UF PUMP VOLUME Confirm Select UF Pump 2 Enter the number of strokes by pressing the and keys Confirm Select pulse amount 60 40 0 10 20 30 Single Needle Test Prime Override Press the Override key Text display for approx 3 sec Confirm Select...

Страница 32: ...I O Bic Conc Set Time Left h min The remaining UF Pulses are shown in the display Confirm Select uf pulses left 59 Ultrafiltration Reset Volume UF Prog I O UF Volume ml UF Rate ml h UF Goal ml Time Left h min 0060 The number of preselected pulses is shown in the UF Goal display Text display after completion Check delivered quantity If necessary adjust UF 2 pump and re peat process Delivery rate 1 ...

Страница 33: ...red Summarized value first measured Refer to the machine card supplied with the machine 2 No more than 1 mA The measurements must be taken in the dialysis mode of operation in the ON phase of the heating control system The scales must be travelled out to such an extent that neither of the two end switches are actuated middle position The two leakage currents are to be measured each time with diffe...

Страница 34: ...and keys Press the Confirm key Brief text display Display on scales is not illuminated Switch off hemodialysis machine Set service switch to ON top Switch on hemodialysis machine Confirm Select CALIBRATION Press the Confirm key Confirm Select CALIB HDF SCALE Confirm Select CHECKING HDF SCALE Text display Confirm Select calib Switch ON Open housing of scales at the back Set calibration switch for t...

Страница 35: ...is active A C appears at the top right 40 0 10 20 30 Single Needle Test Prime Override Press the Override key Text display No weight is to be on the scales Confirm Select HDF SCALE unloaded 40 0 10 20 30 Single Needle Test Prime Override Press the Override key Text display Confirm Select HDF DISPL shows 0 Scales display 50 0 100 g C Fresenius Medical Care ...

Страница 36: ...ext display Scales display 5000 g Hang calibration weight PT6 5000 g on scales Confirm Select load 5 kg CAL WEIGHT 50 0 100 C Fresenius Medical Care 40 0 10 20 30 Single Needle Test Prime Override Press the Override key Text display Confirm Select DISPL shows UNIT g Scales display 5000 g 50 0 100 g Fresenius Medical Care ...

Страница 37: ...ical Care Set calibration switch for the scales to OFF position Close housing 40 0 10 20 30 Single Needle Test Prime Override Press the Override key Text display Remove calibration weight Confirm Select HDF SCALE unloaded 40 0 10 20 30 Single Needle Test Prime Override Press the Override key Brief text display Confirm Select SCALE CALIBRAT done Text display Confirm Select CALIB HDF SCALE ...

Страница 38: ...off hemodialysis machine Set service switch to OFF bottom Switch on hemodialysis machine Hang calibration weight PT6 5000 g on scales Scales display 5000 g 50 0 100 g Fresenius Medical Care Remove calibration weight Calibration of 4008 HDF scales completed ...

Страница 39: ... Medical Care Place filled substituate line in sub stituate sensor Set potentiometer P1 on the LP 754 so that the LED D26 just illuminates XHDF XWAAGE X348V_SN P1 IC14 X348V IC17 IC2 D26 IC8 TP 0 F E D C B A987 6 5 4 3 2 1 I O Spülen Heißreinigen Desinfektion Single Needle Test Vorbereiten Dialyse Start Alarm Ton Aus Bes Luftdetektor Arterieller Druck Venöser Druck TMP Blutleck Überbrücken Ultrafi...

Страница 40: ...volutions the potentiometer is turned Keep this number in your mind 0 F E D C B A987 6 5 4 3 2 1 XHDF XWAAGE X348V_SN P1 IC14 X348V IC17 IC2 D26 IC8 TP Turn back the potentiometer P1 on the LP 754 by half this number The LED D26 will illuminate again XHDF XWAAGE X348V_SN P1 IC14 X348V IC17 IC2 D26 IC8 TP 0 F E D C B A987 6 5 4 3 2 1 ...

Страница 41: ...he hemodialysis machine off Set HEX switch on the LP 754 to position F Switch the hemodialysis machine on Start Stop Rate ml min Ø mm HDF Reset Vol l Start Stop Rate ml min Ø mm HDF Reset Vol l Display 000 Open SN pressure transducer to atmos phere Press the Start Stop key Display 250 Apply a pressure of 250 mmHg to the SN pressure transducer Press the Start Stop key ...

Страница 42: ...3 2 1 X348V_SN Switch off hemodialysis machine Set HEX switch on the LP 754 to position 1 Balancing out has been completed 3 5 2 Checking SN switchover points Bottom switchover point fixed at 75 mmHg 6 mmHg Top switchover points 6 mmHg Stroke volumes ml 10 15 20 25 30 35 40 45 50 Top switchover point 110 130 150 172 195 219 244 270 299 ...

Страница 43: ...re 4008 HDF 3 11 97 TM 3 17 3 6 Repair instructions Important If the substituate lift has to be dis mounted for repair the seal is to be checked and if necessary renewed before fitting the lift again Drip water protection ...

Страница 44: ...3 18 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 45: ...and component layout diagram 4 11 4 2 LP 754 control board HDF 4 15 4 2 1 Description 4 15 4 2 2 Circuit and component layout diagram 4 17 4 3 LP 760 HDF motor control 4 23 4 3 1 Description 4 23 4 3 2 Circuit and component layout diagram 4 25 4 4 LP 761 HDF drive keyboard 4 29 4 4 1 Description 4 29 4 4 2 Circuit and component layout diagram 4 31 4 5 LP 762 scales comm 4 35 4 5 1 Description 4 35...

Страница 46: ...4 2 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 47: ... 4008 4008 E LP 744 power control 1 4008 B M Motor Substituate lift Auxiliary socket X5 X4 X2 X3 X1 LP 760 HDF motor control X2 X1 LP 761 HDF drive keyboard Substituate sensor X2 X1 X3 X4 LP 762 scales comm Module Scales Module Scales Switch Substituate lift LP 763 SEE serial interface extender X637C X637B X637B X637C XA1P XA4V0 UF2 pump V126 LP 747 distribution board X744A X744C Fig Connection di...

Страница 48: ...4 4 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 49: ...9 10 X2 LP 760 HDF motor control X2 X1 LP 761 HDF drive keyboard DGND A1 DGND C1 PWR_GND A28 PWR_GND C28 24V A29 24V C29 12V A31 12V C31 5V A32 5V C32 RXD_EXT_1 C20 TXD_EXT_1 C21 A1 C1 A28 C28 A29 C29 A31 C31 A32 C32 C20 C21 X63Z X63Z LP 630 motherboard LP 763 SEE serial interface extender no assignment 1 no assignment 2 no assignment 3 GND 4 no assignment 5 TXD 6 GND 7 no assignment 8 GND 9 PRINT...

Страница 50: ...4 6 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 51: ... relay M 24 V switch controller Imax 15 A short circuit cut out 12 V switch controller Imax 4 A current limitation LP 630 motherboard LP 633 input board LP 634 output board LP 754 control board HDF LP 760 HDF motor control 24 V 12 V Linear controller current restricted internally to approx 1 A LP 761 HDF drive keyboard 12 V Scales 12 V Main transformer Fig Block circuit diagram mains voltage 4008 ...

Страница 52: ...4 8 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 53: ...nnection to LP 760 Display All necessary information is given to the user through 6 multiplex 7 segment displays and 2 LEDs which are mounted on the socket Keyboard The keyboard comprises 3 single non illuminated keys and 1 key with integrated LED HDF The keys are soldered directly to the P C B The user receives a acknowledgement through a spring contact when the keys are pressed ...

Страница 54: ...4 10 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 55: ...24V X 24V 24V 5V 5V 5V 5V 5V 5V 5V 5V 5V 1 2 3 4 5 6 10 9 8 7 DP1 HDSP7801 1 2 3 4 5 6 10 9 8 7 DP2 HDSP7801 1 2 3 4 5 6 10 9 8 7 DP3 HDSP7801 1 2 3 4 5 6 10 9 8 7 DP4 HDSP7801 1 2 3 4 5 6 10 9 8 7 DP5 HDSP7801 1 2 3 4 5 6 10 9 8 7 DP6 HDSP7801 1 19 2 3 4 5 6 7 8 9 18 17 16 15 14 13 12 11 IC1 74HCT541 8 7 6 5 4 3 2 1 LD1 HLMP2655 1 2 LD2 HLMP2400 3 4 LD2 HLMP2400 R3 4K99 R1 4K99 R6 7K50 R5 7K50 R4...

Страница 56: ...4 12 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 57: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 13 LP 625 Component layout diagram ...

Страница 58: ...4 14 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 59: ...tion Voltage generation The 24 V and 12 V voltage supply is made available to the blood pump by the monitor The 5 V supply voltage is generated on the module by the switch controller IC 20 from the 24 V voltage to minimize loss of power Stepper motor control The stepper motor is run in microstep operation to reduce noise The resolution is 60 microsteps per step The RISC processor transmits a 8 bit...

Страница 60: ...d otherwise the processor would be too slow to generate this frequency RISC processor The RISC processor IC 5 is supplied with a cycle signal from the processor With each cycle an 8 bit data word is read from a look up table by the processor alternately for each phase of the stepper motor respectively Included in this data word is both the current direction as well as the current desired value A w...

Страница 61: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 17 4 2 2 Circuit and component layout diagram LP 754 control board HDF LP 754 Circuit diagram 1 2 ...

Страница 62: ...4 18 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 63: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 19 LP 754 Circuit diagram 2 2 ...

Страница 64: ...4 20 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 65: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 21 LP 754 Component layout diagram ...

Страница 66: ...4 22 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 67: ...the motor a continuous operation R21 with components R22 without components b inching operation for downwards movement R21 without components R22 with components To Point a One D flip flop is available for each direction of movement IC 3 The flip flops are set by a high level on X2 2 key or X2 4 key The flip flop for the respective other direction of movement is reset at the same time by pressing ...

Страница 68: ... caused by the current flow through R 18 is connected to the capacitor C 14 and C 13 through the diodes D 4 and D 5 The resulting direct voltage through R 13 is approx double so high as the peak value of a half wave of the alternate voltage on R 18 less the on state voltages of D 4 and D 5 The direct voltage controls the optocoupler IC 4 through R 14 R 13 serves to completely discharge C 13 and C ...

Страница 69: ...N LC TXD TARA MOTOR_CTRL AGND AGND TARA RXD TXD 12V 12V AGND 12V 12V AGND 12V 12V AGND AGND AGND AGND PE 12V AGND R5 73K2 2 3 4 5 1 IC2 4072 9 10 11 12 13 IC2 4072 1 2 3 T1 7812 C5 100N C4 10U C6 1U0 6 5 3 4 1 2 IC3 4013 8 9 11 10 13 12 IC3 4013 R2 10K0 R3 10K0 R4 10K0 2 X2 3 X2 4 X2 C7 10U C8 1N0 C10 1N0 R1 10K0 6 X1 5 X2 6 X2 7 X2 8 X2 1 X2 2 X3 3 X3 4 X3 5 X3 1 X3 6 X3 4 5 3 1 2 RL1 VSB24 D1 1N...

Страница 70: ...4 26 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 71: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 27 LP 760 Component layout diagram ...

Страница 72: ...4 28 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 73: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 29 4 4 LP 761 HDF drive keyboard 4 4 1 Description General The keys for controlling the substituate lift are found on this P C B ...

Страница 74: ...4 30 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 75: ...O TAG NAME ERSATZ ERSATZ FORMAT BLATT A4 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 MEDIZIN TECHN SYSTEME LP761 673 766 HDF ANTRIEB TASTATUR UP DOWN STOP HDF MOTOR KEYBOARD 1 1 SP 04 08 OTT 1994 05 08 ASCH a X1 geandert 12V UP STOP DOWN SE1 SE2 SE3 SE4 12V SUB_A SUB_K SUB_C PWR_GND 1 2 S2 RF15 1 2 S1 RF15 1 2 S3 RF15 1 X2 2 X2 3 X2 4 X2 5 X2 6 X2 7 X2 8 X2 1 X1 2 X1 3 X1 4 X1 5 X1 6 X1 ...

Страница 76: ...4 32 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 77: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 33 LP 761 Component layout diagram ...

Страница 78: ...4 34 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 79: ...the interface IC 1 max 232 to 5 V level IC 2 makes a serial parallel conversion and passes the signal on to the scales The scales are switched on or off respectively by the transistor T1 The serial carry forward weight comes from the scales to the plug X3 pin 6 TXD and is converted by IC 1 from 5 V to RS 232 level The signal arrives at the dialysis machine through plug X1 pin 5 TXD SP1 ML78L05 con...

Страница 80: ...4 36 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 81: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 37 4 5 2 Circuit and component layout diagram LP 762 scales comm LP 762 Circuit diagram ...

Страница 82: ...4 38 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 83: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 39 LP 762 Component layout diagram ...

Страница 84: ...4 40 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 85: ...P C B Function After switching on the machine the UART is initialled with the transmission parameters Transmission parameter channel A 1200 Baud 8 databits 1 stopbit no paritibit Transmission parameter channel B 9600 Baud 8 databits 1 stopbit no paritibit Channel A is responsible for communication with the scales channel B for communication with the HDF pump module After going through the watchdog...

Страница 86: ...4 42 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 87: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 43 4 6 2 Circuit and component layout diagram LP 763 SSE serial interface extender LP 763 Circuit diagram 1 2 ...

Страница 88: ...4 44 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 89: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 45 LP 763 Circuit diagram 2 2 ...

Страница 90: ...4 46 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 91: ...Fresenius Medical Care 4008 HDF 3 11 97 TM 4 47 LP 763 Component layout diagram ...

Страница 92: ...4 48 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 93: ...l Care 4008 HDF 3 11 97 TM 5 1 Table of contents 5 Spare parts Section Page 5 0 How to use the spare parts catalog 5 3 5 1 P C B s 5 4 5 2 Scales 5 6 5 3 Substituate lift 5 8 5 4 UF2 pump blood pump HDF valve V126 5 10 ...

Страница 94: ...5 2 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Страница 95: ...an assembly can be identified by an increasing decimal e g 5 X 1 5 X 2 In the spare part lists it is refered to Built in From Equipment Code To Equipment Code Update service Updates to this spare parts catalog will be released as Replacement pages Supplementary pages Technical information Generally we reserve the right to make changes Requirements for correct spare part orders The equipment code m...

Страница 96: ...eratur Prog I O Prog I O Bic Konz Set I O mmHg kPa mmHg kPa mmHg kPa ml mS cm 25 C ml min C Set UF Rate ml h UF Ziel ml UF Restzeit h min 4008 Start Stop Rate ml min Ø mm Start Stop Rate ml min Ø mm HDF Reset Vol l Pven Start Stop Rate ml h Bolus ml Bolus Rate h min 40 0 100 200 300 100 200 280 0 10 20 30 10 20 30 0 100 200 0 10 20 30 300 400 500 40 50 60 0 100 200 0 10 20 30 300 400 500 40 50 60 ...

Страница 97: ... Item Part number Description 1 673 790 1 P C B LP 762 scales comm 2 674 182 1 P C B LP 763 SEE serial interface extender 3 673 766 1 P C B LP 761 HDF drive keyboard 4 673 872 1 P C B LP 625 display board 5 673 720 1 P C B LP 760 HDF motor control 6 673 041 1 P C B LP 754 control board HDF ...

Страница 98: ...5 6 Fresenius Medical Care 4008 HDF 3 11 97 TM 5 2 Scales 1 2 3 4 5 6 7 8 19 16 17 18 15 14 13 12 10 9 11 20 21 23 22 24 30 29 28 27 25 26 ...

Страница 99: ...10 673 987 1 Latch 11 642 861 1 Screw 12 553 621 1 Nut 13 673 986 1 Adapter 14 563 151 1 Retaining washer 15 674 106 1 Line holder 16 552 275 1 Screw 17 673 971 1 Fastening plate 18 642 693 1 Screw 19 673 985 1 Bearing 20 672 530 1 Cover 21 552 276 1 Screw 22 674 551 1 Caution label Stability limit 23 673 977 1 Display plate 24 673 922 1 Back shell 25 673 921 1 Front shell 26 553 620 1 Nut 27 559 ...

Страница 100: ...5 8 Fresenius Medical Care 4008 HDF 3 11 97 TM 5 3 Substituate lift 1 2 3 4 5 6 7 8 9 11 14 16 17 6 28 27 25 26 6 24 22 21 20 19 18 35 9 34 33 32 31 30 29 23 10 37 36 38 12 39 40 41 ...

Страница 101: ...18 646 274 1 Tolerance sleeve 19 674 078 1 Stripper 20 673 974 1 Collar bush 21 673 920 1 Housing lift 22 673 639 1 Front foil lift 23 559 690 1 Washer 24 674 400 1 Connection cable 25 674 358 1 Connection cable 26 673 918 1 Base plate 27 673 972 1 Funnel 28 674 255 1 P C B cover 29 646 606 1 Spacer bolt M 3 x 8 5 30 650 439 1 Spacer bolt M 3 x 12 5 31 559 690 1 Washer 32 574 617 1 Spacer sleeve 3...

Страница 102: ... 5 4 UF2 pump blood pump HDF valve V126 Start Stop Rate ml min Ø mm HDF Reset Vol l 40 Hydraulic section 4008 4008 E Hydraulic section 4008 B Concentrate pump 23 UF pump 22 UF2 pump 20 UF2 pump 20 Concentrate pump 23 Bibcarbonate pump 25 1 2 3 4 5 6 ...

Страница 103: ...70 706 1 Ultrafiltration pump complete Blood pump HDF The blood pump HDF is similar to the blood pump single needle see appropriate Spare Parts Catalogue 4008 40 671 726 1 Front foil 673 924 1 Blood pump module HDF complete Valve V126 1 645 881 1 Filter 2 545 361 1 Silicon tube 0 12 m 3 674 349 1 Solenoid valve MV 126 solenoid valve MV43 4 674 347 1 Valve block 5 552 275 1 Screw 6 579 072 1 O ring...

Страница 104: ...5 12 Fresenius Medical Care 4008 HDF 3 11 97 TM ...

Отзывы: