
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 87
F- 7.7.22 LESS THAN 800 ML COLLECTED / ISOLATE VALVE #34
A) Turn the machine OFF!
B) Return Valves #38 AND #31 to their PROPER distribution board positions.
C)
Per the Figure previous page
, replace
Valve #34
with a known good!
D) Plug the new valve #34’s connector into valve #33’s distribution board position, “V33”.
E) ENSURE the bucket with the RED dialyzer connector submersed in it is full!
F) Turn the machine on but
DO NOT
press any other keys!
G) With “Select Program” up, measure drain flow for one (1) minute. More than 800 ml every minute?
Yes
More than 800 ml! Problem located! The previous valve #34 is bad!
No Less than 800 ml! See parts a and b below:
a) Return all valves to their PROPER distribution board positions.
b) THREE (3) possible bad components:
1)
Actuator-Test board;
2)
Actuator cable;
3)
Diistribution board
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...