2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 641
P- 2.0.3 ISOLATE ‘SWITCHED’ 24 V-B
a)
IMPORTANT! TURN THE MACHINE OFF!
b) Figure below, locate the motherboard’s nine (9) pin TEST connector.
Figure 129 – Card Cage / TEST Connector / 24V-B
c) TIGHTLY HOLD the red meter lead on the TEST connector’s pin 8 (2
nd
pin from the right).
d) Looking for
approximately 24.0 volts to momentarily pulse on and off more than once,
WITHOUT
LOOKING AWAY FROM THE METER
, turn the
machine ON (fan running)
and watch for thirty
(30) seconds!
e) FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) momentarily went to between 23.0 and 28.0 volts but after
about thirty (30) seconds dropped to AND remained less than 3.0 volts:
See
2)
IF (and ONLY if) ALWAYS less than 3.0 volts i.e. NEVER more than 23.0 volts!
Perform procedure number P- 2.0.3 once again just to be sure. If still always less than 3.0 volts,
proceed to
page 643
, procedure number
Scenario #3 and #4 next page