2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 618
P- H.1.2 TROUBLESHOOT VOLTAGE DETECTION PROBLEM
Either “Operator Error” occurred OR
5V EST
AND / OR
12V EST
were not in range. These procedures
determines if a component may be ‘dragging’ +12 and / or +5 high or low:
a)
IMPORTANT! To prevent damage turn the machine OFF!
b) Figure below, unplug ALL
MODULES ONLY
, including the Blood Pressure module!
c) From here forward,
i
gnore “Art BP No Comm” OR “Blood Still Sensed!” if they occur.
d) Repeat the Voltage Detection Calibration. THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) “Operator Error” reoccurs:
(page 619).
2) IF (and ONLY if) 5V EST AND / OR 12V EST are not in range:
See procedure number
(page 619).
3)
IF “Operator Error” did NOT occur AND 5V EST AND 12V EST are in range:
TURNING THE
MACHINE OFF IN BETWEEN reconnect the module cables
*
, one at a time, and repeat the
Voltage Detection Calibration in between until the problem reoccurs indicating the last module
plugged may be the problem.
*
CAUTION!
DO NOT plug a module, other than the Blood Pressure module, into the
Blood Pressure or Colin position!
* CAUTION!
DO NOT plug the Blood Pressure module, into another module’s position!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...