F- 1.0.9 SCENARIO #1 / LOADING PRESSURE MORE THAN 29 PSI
Referring to the Figure below, adjust Loading Pressure Valve #65’s nut counterclockwise but no more than a
few turns. Can you adjust PEAK pressure to between 23 and 25 psi (Yes or No)?
Yes
Between 23 and 25 psi! Call debug screen 0. WITHOUT LOOKING AWAY, watch
Flow Error
for five (5) minutes. If (and ONLY if) EVER = 1, even just once, return to (ABOVE) procedure
number
(
page 26
). If
Flow Error
remains = 0 a Flow problem is NOT occurring at
this time but other problems may be.
No REMAINS more than 29 psi! See parts A and B below:
A)
Turn the machine OFF.
B)
Proceed to
page 568
, procedure number
LOADING PRESSURE VALVE #65