FSB# 15-HOM-001 Rev. A   

Page 1 of 7

 

 

 
 
 

Fresenius Medical Care North America 

Service Bulletin 

 

Equipment: 

2008K@Home

 Hemodialysis Machines 

Bulletin: 

15-HOM-001  Rev A  

Subject: 

Optional Shunt Box Ferrite Core Installation (Update) 

 

 

1.0  PURPOSE 

To inform all Facilities, Field Service and Technical Support personnel of how to replace the 
shunt box wire harness (which includes a ferrite core) for ESD related “Dial Valve Failure 1” 
alarms on a 2008K@Home Hemodialysis machine.  This Service Bulletin replaces 
FSB# 08-HOM-002.

 

2.0  PARTS REQUIRED 

 

Figure 1 

– Shunt Switch Assembly (P/N M42408) 

3.0  REQUIRED TOOLS 

  #1 Phillips Screwdriver 

  Masking tape or equivalent 

  #2 Phillips Screwdriver 

  5mm Hex (Allen) Wrench 

4.0  INSTRUCTIONS 

4.1 

Stabilize the machine by applying the caster brake.

 

 

4.2 

Remove any items hanging on the I.V. pole. 

4.3 

Locate and remove the screw attaching the I.V. pole to 
its mounting base using a 5mm Hex (Allen) Wrench.  
See Figure 2 for screw location. 

4.4 

Slide the I.V. pole up and out of the shunt box. 

 

© 2015, Fresenius Medical Care North America, All rights reserved. 

Figure 2 

I.V. Pole Mounting 

Screw Location 

Содержание 2008K Home

Страница 1: ...lysis machine This Service Bulletin replaces FSB 08 HOM 002 2 0 PARTS REQUIRED Figure 1 Shunt Switch Assembly P N M42408 3 0 REQUIRED TOOLS 1 Phillips Screwdriver Masking tape or equivalent 2 Phillips Screwdriver 5mm Hex Allen Wrench 4 0 INSTRUCTIONS 4 1 Stabilize the machine by applying the caster brake 4 2 Remove any items hanging on the I V pole 4 3 Locate and remove the screw attaching the I V...

Страница 2: ...e card cage out far enough to access the four screws indicated by white arrows in Figure 5 and disconnect the shunt box ribbon cable from the back of the card cage 4 8 Remove the four screws securing the shunt box to the cabinet Note One of the four screws secures a plastic clamp used as a cable restraint for the shunt box ribbon cable see Figure 5 Retain this clamp for reinstallation In some case...

Страница 3: ...ons of the shunt box together 4 11 Slide the metal portion out of the plastic portion and locate the pivot shaft ends using the arrows in Figure 7 below Note With the shunt door closed the pivot shaft is held in place by friction Once the shunt door is opened the friction is removed and the shaft can fall out of place Figure 7 Pivot Shaft Location Door Closed In order to prevent the pivot shaft fr...

Страница 4: ...s the shunt switch cover removed Figure 10 Shunt Switch Assembly Uncovered 4 14 The shunt switch assembly is held in place with one screw as illustrated by the arrow in Figure 11 With the shunt door still open remove the screw securing the shunt switch assembly in place Retain this screw and lock washer Note As the shunt switch assembly is being removed make note of how the slit in the shunt switc...

Страница 5: ...box structure using the noted orientation during removal Install the original screw and lock washer but do not tighten the screw Leave the screw loose as shown with the arrow in Figure 13 so the switch can move around during the installation of the switch cover Figure 13 New Shunt Switch Assembly Positioning 4 17 With the shunt door still open refer to Figure 14 to position the shunt switch cover ...

Страница 6: ...15 Figure 15 Shunt Switch Cover Protrusion 4 19 Once the cover is seated properly close the shunt door and align the shunt switch cover over the threaded hole in the shunt switch assembly bracket Install the original screw to secure the cover Figure 16 4 20 Once the shunt switch cover is secured tighten the screw left loose earlier see Figure 17 Figure 17 ...

Страница 7: ...e cabinet and check reseat all connections to the card cage 4 25 Install the screws securing the card cage into the cabinet 4 26 Stow all ribbon cables and wire harnesses and connect the ground wire to the middle panel cover Install the middle rear panel 4 27 Slide the I V pole through the shunt box and into its mounting base Secure the I V pole into place and secure it into place using the origin...

Отзывы: