51
9
Warranty, inspection and repair
9.1
Warranty
9.2
Maintenance requirements
Please refer to the terms and conditions of the local
Fresenius Kabi
organisation or the local distributor for the duration of the warranty.
The warranty covers repair and replacement of components proven to
be defective in material or workmanship. The warranty does not extend
to units that have been altered, repaired or modified by unauthorised
persons and malfunctions which are due to improper handling and wear.
The manufacturer is responsible for the safety, reliability and
performances of the nutrition pump if:
assembly, extensions, readjustments, modifications or repair
have been performed by persons authorised by
Fresenius Kabi,
the electrical installation site where the pump is used complies
with the requirements of the IEC regulations,
the pump is used as specified in these Instructions for use,
according to its intended use as defined in Section 1.1, page 6
the pump is operated with the sets specified by the manufacturer
(see Section 2.3, page 15).
The recommended maintenance interval for the
Ambix
activ
and the
pump holder is two years or the application of 1,000 litres (equivalent to
8,000 hours at 125 ml/h). The
Ambix
activ
indicates the application of
1,000 litres by showing the wrench symbol permanently in the display.
Pump performance over life cycle is ensured only if the recommended
maintenance is performed.
Warning
Pump and pump holder may only be repaired by the
manufacturer’s Pump Maintenance Department or persons
authorised by them. Failure to observe this leads to loss of
warranty. In the event of a fault, always send in the complete
system (pump, pump holder and set).
If the device is modified (repaired), appropriate inspection and
testing must be conducted to ensure continuated safe use of
equipment.