![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 275](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626275.webp)
127
18.8
Dimensions and Weight
18.9
Trumpet and Start-up Curves
H / W / D
135 x 345 x 170 mm
Weight
Approximately 2.1 kg
Screen Size
70 x 35 mm
The trumpet curve shows the variation of the mean flow rate accuracy
over specific observation periods. The variations are presented only as
maximum and minimum deviations from the overall mean flow within the
observation window.
Trumpet curves are presented below for a number of representative flow
rates.
The test protocol used to obtain these results is described in ANSI/AAMI
ID26.
The curves can be helpful in determining the suitability of infusion
parameters for specific drugs and concentrations.
Syringe used
: BD Precise 50 mL
Fluid used
: distilled water
Recommendations to improve performances and safety when the pump
is commonly used at low flow rates (
≤
20 mL/h):
Limit the range of available flow rates in accordance with the
maximum flow rate to be used.
Lower the pressure limit in order to gain in time to detect
occlusion.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...