Power Management
17.1
AC Power Supply Precautions
17.2
Battery Precautions
Check that the AC power supply voltage corresponds to the value
indicated on the label on the bottom of the device. Do not exceed the
permitted voltage.
The power outlet must remain accessible at all times to allow emergency
power supply disconnection.
Warning
The pump and its accessories can only be connected to the AC
power supply with the power cord supplied by
Fresenius Kabi
, or
with a power supply accessory from the Agilia product range.
Do not use an extension cord when connecting the pump to the
AC power supply.
Pumps must be plugged into a medical grade power strip if one is
used.
The device uses a Lithium-ion rechargeable battery.
The following actions may cause leakage, overheating, smoke,
explosion or fire; which could result in deterioration of performance,
failure, damage to the equipment or injury to the user:
Incorrect handling of a Lithium-ion battery.
Replacement of the battery by inadequately trained personnel.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...