![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 266](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626266.webp)
118
16
Cleaning and Disinfecting
16.1
When to Clean and Disinfect the Pump
To avoid the risks of infection and microbial transmission, make sure to
adequately clean and disinfect the equipment.
Warning
The disinfecting procedure must be done immediately after
cleaning. Disinfecting the pump without prior cleaning is not
effective.
The pump is not intended to be sterilized; sterilization may result
in damage to the pump.
In case of contamination by blood or bodily fluids when the pump
is in use, and if allowed by your local practices and healthcare
facility policies, immediately perform the quick cleaning described
below. Always follow your local protection rules.
Quick Cleaning Only
Note: This quick cleaning does not replace the need for a complete
cleaning.
1.
Check that the keypad is locked in order to avoid unintended
modification of the infusion parameters. Do not move the pump.
2.
Use ready-to-use wipes to wipe down all exposed surfaces of the
pump.
3.
At the end of the infusion, perform the complete cleaning protocol,
Thoroughly clean and disinfect the pump in the following cases:
After each patient use
Before any maintenance
On a routine basis when the pump is not in use
Before storage
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...