115
15.9.4
Volume Accuracy
* Test condition:
Back pressure:
0 mmHg
**
Syringe
: BD Precise
A back flow pumping is provided to reduce the bolus volume at occlusion release.
During pump movement from 0 to 1 m above the patient, a bolus (-0.05
≤
X
≤
0.35 mL) may occur.
15.9.5
Pressure Accuracy
Warning
Accuracy may be reduced when the infusion flow rate is below 1 mL/h.
Accuracy
Direct Bolus*
≤
10 mL: ± 0.2 mL
> 10 mL: ± 3%
Programmed Bolus*
Bolus Volume at Occlusion
Release*
Syringe**
Rate
Bolus Volume at Occlusion Release
50 mmHg
900 mmHg
50 mL
5 mL/h
-0.05
≤
X
≤
0.35 mL
-0.05
≤
X
≤
0.35 mL
Accuracy
Pressure
< 500 mmHg: ± 75 mmHg
> 500 mmHg: ± 15 %