92
10
Data Communication
10.1
Overview
10.2
Communication via Agilia Cables
10.2.1
Data Communication Cables
Cable Communication
Wi-Fi Communication
Connection of 1 pump to a PC for the following
purposes:
Data set upload (via Agilia Vigilant Drug’Lib
software)
Maintenance (via Agilia Partner software)
Communication between a hospital information
system server and a number of identified pumps
for the following purposes:
Data set upload
Pump history retrieval
Information
Ensure that all hospital information systems have been approved
by
Fresenius Kabi
. For more information, contact your technical
services representative.
Before connecting the pump to a hospital information system, ask
your IT or biomedical department to configure the device.
Information
Only use recommended Agilia cables.
All connections and disconnections must be performed by
qualified and appropriately trained staff.
All IT devices (including computers, hubs and switches) inside the
patient area (< 1.5 m) must comply with IEC/EN 60601-1
(leakage current).
IT devices connected outside the patient area (> 1.5 m) must be
at least IEC/EN 60950 compliant.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...