![Fresenius Kabi Agilia SP Скачать руководство пользователя страница 132](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626132.webp)
132
12
Technical Characteristics
12.1
Power Supply
12.2
Battery
= Wi-Fi enabled
= Wi-Fi disabled or not used
12.3
Power Consumption
Use the power supply cord supplied with the pump.
For use in the US, it is mandatory to use an Agilia power cord compliant with USA standards and with the
IEC 60227 standard.
Maximum admissible apparent electrical cable network impedance: 50
Ω
.
The power cord conductor must have a cross section of at least 0.75 mm
2
.
AC Power
Power supply
100 V - 240 V ~ / 50 / 60 Hz with functional earth
Maximum consumption
10-15 VA
Protective fuse
1 X T1.6AH 250V accessible in the battery
compartment
External
Power Supply
(Available in a
Future Release)
9 VDC / Power > 15 W
Cord
Length: 3.00 m
Maximum length: 3.00 m
Type: SJT3x18AWG
Technical data: 3x18AWG (Green, black, white), dielectric strength: 2 kVac, 1 min
Standard: UL62 CSA 22.2 No. 49
Plug
Type: HG/TR
Technical data: 15A - 12.5V AC.
Standard: NEMA 5-15 Hospital Grade
Connector
Type: C13
Technical data: 10A - 250V UL CSA
Standard: EN 60320-1 C22.2 N°21 UL 817
Disconnect the battery before opening the device. Avoid short circuits and extreme temperatures.
If the device is not used for more than 3 months, the date is erased (all other settings are stored
permanently). When you power on the pump, you must set the date again.
Characteristics
7.2 V 2.2 Ah - Li-ion Smart battery
Weight
Approximately 100 g
Battery Life
Flow Rate
Wi-Fi
Battery Life
5 mL/h
5 mL/h
> 9 h
> 13 h
Battery Recharge
Pump OFF: < 6 h / Pump ON: < 20 h
The infusion pump typically consumes about 3.5 W in standard operating conditions.
Содержание Agilia SP
Страница 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Страница 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Страница 145: ...145 W Watchdog 58 Wi Fi Board Description 24 Dismantle Re assemble 91 Wi Fi Module Information 59 ...
Страница 147: ...Local Contacts for Servicing Write your contacts here ...
Страница 148: ...11378 2_tm_agilia_sp_eng ...
Страница 305: ...Local Contacts for Servicing Write your contacts here ...
Страница 306: ...11103 2_master_ifu_agilia_sp_mc_eng ...