7.9
Patient
7.10
Day/Night Mode
Symbol
Procedure
Changing a patient’s weight or body surface area
Information
If the selected dose rate unit is weight-based (kg), the screen
displays the patient’s weight.
If the selected dose rate unit is body surface area-based (m²), the
screen displays the patient’s body surface area.
You can change the patient’s weight or body surface area as follows:
1.
Press
.
2.
Press the arrow keys to select
.
3.
Press
enter
.
4.
Press
OK
to change the patient’s weight or body surface area.
5.
Press
OK
to confirm the infusion settings.
Symbol
Procedure
Switching between day mode and night mode
This function switches between day mode
and night mode
.
The default night mode settings are as follows:
The key-press beep is silenced.
Infusion indicators and screen brightness are dimmed.
Depending on your pump configuration, the switch between day and night mode
may be managed either through this menu (manual mode), or according to pre-
defined settings (auto mode). For more information, refer to the technical manual.
Weight
Body Surface Area