6.10.3 Volume Limit
You can use this infusion mode to set a limit to the volume infused. When the limit
is reached, an alarm is triggered.
1.
Access the
Volume Limit
menu, see section 7.12, page 66.
2.
Press the arrow keys to set the volume limit, and press
OK
.
Information
If you program a volume limit that exceeds the actual volume in
the syringe, make sure to replace the syringe when it is empty,
The volume already infused (VI) before accessing the volume limit
mode is taken into account.
3.
Press the arrow keys to configure the end of infusion settings and press
OK
to
confirm.
Stop:
The infusion stops when the volume limit is reached.
Keep Vein Open (KVO):
After the volume limit is reached, the infusion
continues at a preset flow rate to keep the access device open.
Continuous:
After the volume limit is reached, the infusion continues at the
programmed flow rate.
4.
Press
start
to start the infusion.
Flow rate
Dose Rate
Flow rate
Dose Rate