FRC DATATRON FTC200 Скачать руководство пользователя страница 1

FTC200 Rev151207

1

Document Number:

XE-FTC2PM-R0A

PORTABLE SAMPLING

FLOW & PRESSURE METER

MODEL FTC200

FIRE RESEARCH CORPORATION

www.fireresearch.com

26 Southern Blvd., Nesconset, NY 11767

TEL  631.724.8888   FAX  631.360.9727   TOLL FREE  1.800.645.0074

Содержание DATATRON FTC200

Страница 1: ...cument Number XE FTC2PM R0A PORTABLE SAMPLING FLOW PRESSURE METER MODEL FTC200 FIRE RESEARCH CORPORATION www fireresearch com 26 Southern Blvd Nesconset NY 11767 TEL 631 724 8888 FAX 631 360 9727 TOLL...

Страница 2: ...nagement System 9 Making Connections 10 Channels 10 Recording 11 Data Management 12 PROGRAMMING 14 Enter Password for Program Access 14 Operator P 100 Codes 16 Set Date and Time codes P107 and P108 17...

Страница 3: ...rror Codes and Fault Warnings Troubleshooting 22 Table 3 Parts List 23 Table 4 Pressure Sensor Output Voltage 26 List of Figures Figure 1 Controls and Indicators 5 Figure 2 Sample File Format 12 Figur...

Страница 4: ...le The FTC200 has a 4 digit Red LED display with daylight bright digits 0 39 inch 10 mm high for PRESSURE readings a5 digitBlueLEDdisplayforFLOW anda2 digitLEDdisplayforbothTIMEandCOUNTS Theflowmeter...

Страница 5: ...FTC200 Rev151207 5 Figure 1 Controls and Indicators Panel Diagram Display Module SYSTEM COMPONENTS...

Страница 6: ...2 DIGIT TIME DISPLAY 2 DIGIT COUNTS Set recording parameters quickly easily BUTTON AUTO Press to automatically sample all parameters on a regular basis BUTTON MANUAL Press to take a single snap shot...

Страница 7: ...be used to power and charge the DataTron Plug the provided DC Power Cable into the POWER input on the front panel and connect the other end to a DC source Jumper Cable The jumper cable is provided to...

Страница 8: ...ve compliant with USB 2 0 and formatted to the standard FAT32 file system for data retrieval by inserting into the MEMORY input on the right side of the front panel of the DataTron Press the MEMORY bu...

Страница 9: ...d in the DataTron is a maintenance free LiFe PO4 non toxic recyclable lead free When the battery is almost depleted use either the external power supply for a wall outlet accessory adapter or jumper c...

Страница 10: ...ow Tube Control 1 To verify pipe size press and hold any flow channel button 1 2 3 or 4 and before the button is released the FLOW displayshowsthepipesizeforthatchannelineithermillimetersorinches Afte...

Страница 11: ...nce the values have been set they will be saved in memory until the next time it is turned back on The back up memory retains the values that were preset last even when the unit is powered off NOTE TI...

Страница 12: ...e DataTron is currently performing another action The current action must either be completed or disabled before a new action can be initiated For example if the user is attempting to transfer data to...

Страница 13: ...First close any open files B Then safely remove the external drive through your computer software For example Windows users will see the safely remove hardware icon C Press the TO COMPUTER button and...

Страница 14: ...Tron is currently performing another action The current action must either be completed or disabled before a new action can be initiated For example if the user is attempting to transfer data to a USB...

Страница 15: ...ord Restricted Contact FRC Provides access to system programming P200 and P300 codes Password Restricted OEM Dealer Programs For more information refer to the OEM Dealer P200 and P300 Codes listed in...

Страница 16: ...ons Refer to details on pages P101 Displays System Model Number FTCXXX Read Only n a P102 Displays Software Revision r1 00 Read Only n a P103 Displays Manufacturing Date 2014 12 08 Read Only n a P104...

Страница 17: ...left side of the FLOW display e g Y 2014 month is symbolized with a lowercase n and d for the day When adjusting the time note the device runs on a 24 hour clock and that minutes are represented by a...

Страница 18: ...ur USB flash drive into the MEMORY port on the DataTron 5 The COUNTS display will read as no 6 Use the UP and DOWN arrow buttons to change the setting to y YES 7 Press and hold the MENU button for app...

Страница 19: ...is allows the flow tube to be used in any flow pressure channel of the currently used device or another DataTron Single Point Calibration 1 First be certain to connect the flow tube into the desired f...

Страница 20: ...nter the new or corrected flow value See note 1 on page 19 7 Editthevalueandoncethecalibrationpointhasbeenenteredintheflowdisplay pressandimmediatelyrelease the MENU button The COUNTS display will adv...

Страница 21: ...another DataTron 7 There are always two calibration points for Pressure Calibration one lower and one upper value Pressure Calibration Steps 1 First be certain to connect the flow tube into the desir...

Страница 22: ...eading a value too close to another calibration point See detailed information for multiple point flow calibration E205 The minimum number of calibration points has not been reached Pressure requires...

Страница 23: ...BA XE XX552 RPA R Adapter AC 12V 2 5A 30W Wall Adapter XE XXPLUGDEU2P A0A Seal Plug 2 Pin for POWER port XE XX6PDP A0A Seal Plug 6 Pin for Flow Tube channels Accessory Plug Part No XE XXBC ZA5074 B Ca...

Страница 24: ...rs from full charge Typical Battery Charge Time 4 5 hours Flow Tubes Material Aluminum Sizes Available 1 5 Tube w 2 5 Couplings 13 320 GPM 2 0 Tube w 2 5 Couplings 21 520 GPM 2 5 Tube w 2 5 Couplings...

Страница 25: ...3 Flow Sensor Wiring Flow Sensor Sensor Cable 3 Pin Connector Pin Wire Description A Black Ground B Red Supply Voltage C White Signal Signal Output Supply Voltage Ground Sensor Cable from 8 Pin Conne...

Страница 26: ...Proof Pressure 800 PSI Type Stainless Steel Excitation Voltage 5 VDC Output Voltage 0 5 VDC 4 75 VDC See Table 1 Table 4 Pressure Sensor Output Voltage 0 psi 100 psi 150 psi 200 psi 250 psi 300 psi 6...

Страница 27: ...r Wiring Pressure Sensor Side View Pressure Sensor Cable 3 Pin Connector Pin Wire Description A Black Ground B Red Supply Voltage C White Signal Pressure Sensor Pressure Sensor Top View Pressure Senso...

Страница 28: ...Remove retaining nut 2 Slide flow sensor out of sensor housing Install Flow Sensor 1 Insert flow sensor into sensor housing Align flat spot on sensor rim with alignment tab and make sure O ring is in...

Страница 29: ...FTC200 Rev151207 29 NOTES...

Страница 30: ...FTC200 Rev151207 30 NOTES...

Страница 31: ...FTC200 Rev151207 31 NOTES...

Страница 32: ...FTC200 Rev151207 32...