7- Fault Finding
All fault finding must be conducted by a competent person or qualified Electrical Engineer.
Please refer to the following table for the trouble-shooting of Xenon lamps.
Fault
Cause
Remedy
◼
Wrong Polarity
◼
Lamp incorrectly fitted
◼
Faulty wiring
◼
Anode (large electrode)
must always be on top in
vertical burning position
◼
Check polarity, transpose
connections if necessary
◼
Cap overheated
◼
Cap temperature above
230°C
◼
Faulty contacts
◼
Cooling equipment
defective
◼
Check terminals, tighten or
renew
◼
Check cooling equipment
and replace if necessary
◼
Arc unsteady
◼
Lamp operated outside
current control range
◼
Magnetic stabilisation for
horizontal operation
defective
◼
Correct current setting
◼
Check magnetic
stabilisation
◼
Bulb draws in air
◼
Crack in graded seal
caused by overheated cap
◼
Maximum cap temperature
230°C
◼
Check terminals - tighten
or renew
◼
Glass erosion on fused
quartz bulb
◼
Lamp operated outside
current control range
◼
Lamp service life
exceeded
◼
Correct current setting
◼
Check meter
◼
Electrodes damaged
◼
Premature blackening
◼
Current ripple too high
◼
Auxiliary mirror incorrectly
adjusted
◼
Have power supply
inspected
◼
Adjust auxiliary mirror
◼
Asymmetrical blackening
of lamp (in horizontal
burning position)
◼
Lamp operated too long in
same position
◼
Turn lamp through 180°
after half service life