User Manual 6008 1286 / Rev. 8
6:3
6.3
Error Messages
Fault
Action
1
Too much light - Sample too
light
Check sample cup
2
Too little light - Sample too dark Check sample cup
3
Internal error: Inconsistent
measurements
Please restart instrument. Contact your FOSS
representative if problem persists
4
Internal error: Invalid
measurement settings
Contact your network administrator for a
new update session
5
Instrument temperature is too
low
Re-start instrument. If problem persists
contact FOSS
6
Instrument temperature is too
high
Turn off instrument for minumum 15 minutes
before re-starting. Contact your FOSS
representative if problem persists
7
Lamp voltage is too low
Contact your FOSS representative
8
Lamp voltage is too high
Contact your FOSS representative
9
Lamp current is too low
Replace lamp. Contact your FOSS
representative if problem persists
10
Lamp current is too high
Contact your FOSS representative
11
Sample cup cannot find start
position
Check if sample cup is jammed
12
Sample cup motor cannot
maintain speed
Check if sample cup is jammed
13
Sample cup has slipped during
measurement
Check if sample cup is jammed
14
Spectrometer communication
error
Contact your FOSS representative
15
Problem in light path
Contact your FOSS representative
16
Instrument is in the process of
stabilising
Wait a minute
17
Instrument unable to keep
detectors at required
temperature
Turn off instrument for minumum 15 minutes
before re-starting. Contact your FOSS
representative if problem persists
18
Reference paddle cannot find
start position
Contact your FOSS representative
19
Reference paddle motor cannot
maintain speed
Contact your FOSS representative
20
Reference paddle has moved
during measurement
Contact your FOSS representative
21
Reference paddle is possibly
blocked
Contact your FOSS representative
22
Cannot communicate with RFID
reader
Contact your FOSS representative
23
Cannot read RFID chip on
sample cup
Check sample cup
24
DDA has not been calibrated
Contact your FOSS representative
Содержание NIRS DA1650
Страница 20: ...NIRS DA1650 2 12 User Manual 6008 1286 Rev 8 ...
Страница 26: ...NIRS DA1650 3 6 User Manual 6008 1286 Rev 8 Fig 3 7 9 Press Install to start installation ...
Страница 30: ...NIRS DA1650 3 10 User Manual 6008 1286 Rev 8 ...
Страница 44: ...NIRS DA1650 4 14 User Manual 6008 1286 Rev 8 Fig 4 25 Fig 4 26 Fig 4 27 ...
Страница 64: ...NIRS DA1650 5 4 User Manual 6008 1286 Rev 8 ...
Страница 68: ...NIRS DA1650 6 4 User Manual 6008 1286 Rev 8 ...
Страница 69: ...NIRS DA1650 User Manual 6008 1286 Rev 8 7 1 7 Parts Accessories and Consumables See the Owner s Guide ...