background image

  

UV-4 Flame Detector 
372000-28 Rev J 

 

 

Section 5 Operation 

The Forney UV-4 detector normally is sighted on the primary combustion zone of a burner, and light from 

the burner flame passes through a sight tube or a fiber-optic light guide to an opening in the base assembly 

housing. Because the sensor tube is positioned directly behind this opening, essentially all light enters the 

tube. IF the intensity of the UV light is great enough to ionize the gas in the tube, the flame detector head 

generates a continuous pulse train. 

5.1  Normal Operation 

Electronic components in the UV-4 flame detector PCB include a dc-dc voltage converter, RC biasing 

network, optocoupler, and single-shot. The external amplifier pr14 vdc power, which is applied to 

all active components on the PCB. The dc-dc voltage converter boosts the input power to produce a large 

potential difference between the anode and cathode inside the UV sensor tube. The sensor tube contains a 

gas that can be ionized by absorbing energy directly from UV radiation. If the rate of ionization exceeds the 

rate of recombination, a surge of current passes through the tube, producing a high-voltage transient. The 

tube operates with a nominal 350-vdc bias voltage and requires a threshold flux of UV radiation to start 

conducting. Each time the tube enters the conducting state, a capacitor in the anode circuitry discharges 

through the cathode circuit, dropping the bias below 350 vdc. This drop in bias voltage causes the tube to 

stop conducting almost immediately, and the power supply begins recharging the capacitor in the anode 

circuit. Since the RC time constants for the anode and cathode circuits are fixed, the interval between 

successive discharges is directly proportional to the magnitude of UV flux incident on the tube. 

 

Each discharge through the sensor tube produces a high-voltage pulse having a duration of no more than a 

few microseconds. Since the magnitude of the voltage pulse is far too great for logic circuitry, the 

optocoupler performs level conversion and also inverts the signal to produce a low-active trigger pulse. The 

leading edge of each trigger pulse then causes the single-shot to generate a digital output pulse having a 

fixed pulse width of 500 

µ

s. When considered as a function of time, the frequency of the output signal from 

the head is directly proportional to the intensity of UV radiant energy striking the tube. 

5.2  Self-Test Operation 

The external amplifier normally runs a self-test cycle at regular intervals under program control. At the start 

of each self-test cycle, the amplifier e15 vdc to energize the shutter solenoid, blocking the light path 

mechanically. The amplifier then monitors the input signal from the detector head to verify that a flame-out 

condition is detected. If the input signal continues to indicate flame presence for the entire test interval, the 

amplifier activates an alarm to indicate test failure. 

Section 6 Maintenance 

The only maintenance needed by the UV-4 detector head is periodic cleaning of the glass lens (direct-view 

configuration only). The appropriate cleaning schedule is determined by the quality and cleanliness of the 

cooling-/seal-air source. Forney recommends the use of a non-oily cleaning solution and a soft lint-free cloth 

for lens cleaning. 

Содержание UV-4

Страница 1: ...UV 4 Flame Detector 372000 28 Rev J ...

Страница 2: ...OTICE The contents of this publication are proprietary data of Forney Corporation Forney reserves all proprietary rights to this document including all design manufacturing reproduction use and sales rights thereto except to the extent said rights are expressly granted to others SAFETY ICON DEFINITIONS DANGER Indicates a hazardous situation which if not avoided will result in death or serious inju...

Страница 3: ... with Fiber Optics Light Guide 3 1 1 Mechanical Specifications 4 1 2 Electrical Specifications 4 Section 2 Safety 5 Figure 5 Cable Connections to Detector Head 5 Section 3 Commissioning 6 Section 4 Removal Installation 6 Section 5 Operation 7 5 1 Normal Operation 7 5 2 Self Test Operation 7 Section 6 Maintenance 7 Section 7 Troubleshooting 8 Table 1 UV 4 Flame Detector Troubleshooting 8 Section 8 ...

Страница 4: ...bassemblies a head assembly and a base assembly The head assembly figure 1 consists of a machined housing that contains the internal electronics of the flame detector Figure 1 UV 4 Head Configuration The sensor tube contains an ultraviolet sensitive gas at reduced pressure When UV light enters the tube the gas ionizes enabling current to flow between two electrodes Signal conversion electronics on...

Страница 5: ...es the electronic components from furnace gases in the sight tube The flame detector is available with either pigtail wiring or a bulkhead connector that provides the electrical interface with the external amplifier assembly Figure 2 Base Assembly with Bulkhead Connector Figure 3 Base Assemblies with Pigtail Wiring A base with a fiber optic light guide This assembly includes a fiber optic cable as...

Страница 6: ...UV 4 Flame Detector 372000 28 Rev J 3 Figure 4 Base Assemblies with Fiber Optics Light Guide ...

Страница 7: ...ture Refer to submittal drawings for figures specific to this project Spectral Response UV mode 190 to 260 nanometers 1 2 Electrical Specifications Shutter Solenoid Power 14 1 vdc 0 9 to 1 0 A nominal current Input Power _15 vdc 47 mA nominal current Output Signal Open collector 500 µs digital pulse train Power Consumption 12 vA maximum I O Cable Connector MS3116E10 6S units with bulkhead connecto...

Страница 8: ...se the ball valve attached to the sight tube In addition maintenance personnel should wear safety glasses to prevent eye injury from pressurized gas and flying particles CAUTION The detector head contains sensitive electrical components Ensure that all power is removed from the external amplifier assembly before connecting or disconnecting the I O cable NOTICE Disconnect the cable from the detecto...

Страница 9: ...base assemblies from the furnace mounting b Reassemble the flame detector and place a UV light source 1 to 4 feet from the surface of the lens or the end of the fiber optic light guide c Use an oscilloscope to measure the signal from the flame detector head at the input terminal of the external amplifier The signal observed should be a series of digital pulses having a maximum frequency between 1 ...

Страница 10: ...egins recharging the capacitor in the anode circuit Since the RC time constants for the anode and cathode circuits are fixed the interval between successive discharges is directly proportional to the magnitude of UV flux incident on the tube Each discharge through the sensor tube produces a high voltage pulse having a duration of no more than a few microseconds Since the magnitude of the voltage p...

Страница 11: ...avoid scratching the glass 3 Replace detector head Return unit for repair or replacement No signal output 1 Supply voltage to head not present 2 Detector head assembly not functional 3 Overheated detector head 1 Verify that external amplifier is supplying the correct voltages for the selected operating mode 2 Replace detector 3 Check the surface temperature of the detector If the temperature is gr...

Страница 12: ...y material to Forney please contact your Forney customer service representative and provide the contract number or the customer purchase order number Contact Forney s Aftermarket Department via any one of the following methods and furnish the following information E mail Phone Fax spares forneycorp com 972 458 6100 or 972 458 6142 or 1 800 356 7740 24 hour direct line 972 458 6600 1 Contract numbe...

Отзывы: