
Troubleshooting |
37
Error
Cause
Solution
The device state is currently
invalid
The Form Wash L attempted
to cancel an ongoing cycle
while the machine was in an
invalid state.
Disconnect and reconnect the power.
An unknown error occurred
Other/unspecified cause
Disconnect and reconnect the power.
Any other error related to
the solvent monitor
The Form Wash L detected an
issue with the solvent monitor
•
Check that the solvent monitor is
submerged in the solvent
•
Check that the solvent monitor
cable is fully plugged into the Form
Wash L
•
Check that the selected solvent
matches the solvent in the Form
Wash L
•
Replace the solvent monitor
Содержание Form Wash L
Страница 1: ...Manual Form Wash L...
Страница 2: ......
Страница 6: ......
Страница 46: ...Original English instructions November 2021 Formlabs...