Troubleshooting |
35
7.3.2
Tripropylene glycol monomethyl ether (TPM)
Spilled TPM does not present an immediate fire or health hazard, but it does leave an oily
residue that can be slippery. TPM evaporates very slowly, so the spill won’t dry by itself.
To clean a TPM spill:
1. Put on a pair of disposable nitrile gloves.
2. If a wash cycle is in progress, cancel the wash on the touchscreen.
3. Unplug the device from its power source.
4. Absorb excess TPM with paper towels.
5. Clean up any remaining residue using wet paper towels.
6. Dispose of tools used to clean up the spill (paper towels soaked with TPM, gloves coated
in TPM) with hazardous waste according to local regulations.
7.4
Troubleshooting errors or abnormal activity
In the case of errors or abnormal activity with the Form Wash L, reference the following
errors, causes, and proposed solutions. Complete the initial troubleshooting steps and
carefully document all results. Contact
additional guidance.
7.4.1
Resolving abnormal functions
Error
Cause
Solution
The display does not
turn on.
Power failure or a faulty
electrical connection
•
Disconnect and reconnect
the power.
•
Plug the power cable into a
different outlet.
The touchscreen
is unresponsive.
Faulty or damaged touchscreen
•
Disconnect and reconnect
the power.
•
Replace the touchscreen.
The outer lid does not
close fully.
Printed part is obstructing
outer lid
Outer lid hinge is broken
Check that any printed parts in
the basket do not interfere with
the outer lid. If the outer lid is
free of any obstructions, contact
The impeller does
not rotate.
•
Debris or obstructions in the
wash bucket
•
Faulty or damaged impeller
motor
•
Check for loose cured material
or debris in the bottom of the
wash bucket.
•
Disconnect and reconnect
the power.
Unusual noises
•
Debris or obstructions in the
wash bucket
•
Improper wash bucket
alignment
•
Check for loose cured material
or debris in the bottom of the
wash bucket.
•
Reseat the wash bucket
Washed parts have tacky or
sticky surfaces.
Resin concentration in solvent
is too high
Replace the solvent. Refer to section
6.4.1 Replacing solvent
.
The machine does not
detect the wash bucket.
Wash bucket not fully inserted
Reseat the wash bucket
Содержание Form Wash L
Страница 1: ...Manual Form Wash L...
Страница 2: ......
Страница 6: ......
Страница 46: ...Original English instructions November 2021 Formlabs...