Error
Cause
Solution
The heaters do not reach
the target temperature.
•
Abnormal display behavior
•
Environmental conditions
•
Faulty or damaged heater
•
Disconnect and reconnect the power.
•
Ensure that the operating
environment is in the recommended
temperature range.
•
Check that the heater fan is spinning.
•
Check and compare the internal
temperature readings from both
heater modules.
The turntable does not turn.
•
Turntable is obstructed
•
Turntable is not fully seated
•
Turntable gear has
separated from the
turntable
•
Faulty or damaged motor
assembly
•
Ensure that no cured resin or printed
parts are blocking the turntable.
Reorient large prints as necessary.
•
Reseat the turntable on its mount.
•
Check that the gear on the bottom
of the turntable is firmly adhered to
the turntable.
•
Check that the rollers underneath the
turntable turn smoothly.
•
Replace the motor assembly.
Parts are undercured
or do not have desired
mechanical properties.
•
Expired resin
•
Part was washed but not
fully dried before post-
curing
•
Check the expected lifetime of the
resin used for the print.
•
Fully dry all solvent off of parts before
post-curing.
Post-cured parts have tacky
or sticky surfaces.
•
Part was not washed
before post-curing
•
Part was washed but not
fully dried before post-
curing
•
Wash liquid resin off of printed parts
before postcuring.
•
Fully dry all solvent off of parts before
post-curing.
7.3.2
Resolving errors
If errors persist after following these steps, contact
or a
Error
Cause
Solution
n
The device is
currently active.
The Form Cure L attempted
to start a post-curing cycle
while a cycle was already
in progress.
Disconnect and reconnect the power.
The device state is
currently invalid.
Firmware error
Disconnect and reconnect the power.
The device’s duty cycle is
currently invalid.
Firmware error
Disconnect and reconnect the power.
The call to DBus failed.
Firmware error
Disconnect and reconnect the power.
The reply to the DBus
was invalid.
Firmware error
Disconnect and reconnect the power.
The fan control failed.
Firmware error
Disconnect and reconnect the power.
The timer has expired.
The Form Cure L is not
receiving temperature data.
Disconnect and reconnect the power.