Carefully examine the product for
damage, missing parts or foreign
objects. Remove any foreign objects
prior to use. The product should be
replaced IMMEDIATELY if there are
any damaged or missing parts. If
necessary, use the Metered Dose
Inhaler (MDI) alone until a
replacement is obtained. If the
patient's symptoms worsen, please
seek immediate medical attention.

Remove cap(s)
from the MDI
and chamber
(if applicable).

Shake the MDI
immediately
before each
use as per the
instructions
supplied with
the MDI.

Insert the
MDI into the
backpiece of
the chamber.

Follow
instructions
supplied with
the MDI on
how long to
wait before
repeating
steps 3 – 6 as
prescribed.

Put mouthpiece into mouth and close
lips around it to ensure an effective
seal. The

Flow-Vu

® Inspiratory Flow

Indicator only moves if the patient
has a good seal.

INDICATIONS FOR USE

This product is intended to be used by patients

who are under the care or treatment of a

physician or licensed healthcare professional.

The device is intended to be used by these

patients to administer aerosolized medication

from most pressurized Metered Dose Inhalers.

The intended environments for use include the

home, hospitals and clinics.

MOUTHPIECE

LARGE MASK

LA

R

G

E

M

A

S

K

M

O

U

T

H

P

IE

C

E

Breathe out gently and depress the MDI at the beginning of a slow inhalation. Use the

Flow-Vu

® Indicator to assist in the coordination of this maneuver. Breathe in slowly and

deeply through the mouth until a full breath has been taken. Hold breath for 5 – 10 seconds,
if possible. Otherwise, keep lips tight on the mouthpiece to maintain seal for 2 – 3 breaths
after the MDI is depressed.

Slow down inhalation if you hear the

FlowSignal

®

Whistle

sound.

Administer one (1) puff at a time.

INSTRUCTIONS FOR USE

THIS PRODUCT CAN BE USED RIGHT OUT OF PACKAGE. BEFORE USE, ENSURE THESE INSTRUCTIONS AND THE INSTRUCTIONS SUPPLIED WITH THE METERED DOSE INHALER (MDI) HAVE BEEN READ.

1

Remove the Backpiece only. Do not tamper
with valve during cleaning or disassemble
the product beyond what is recommended
or damage may result. For mask product,
do not remove mask.

Soak the parts for 15 minutes in a mild solution
of liquid dish detergent and lukewarm clean
water. Agitate gently.

Rinse parts in clean water.

Shake out excess water from the parts
and allow to air dry in a vertical position.
Ensure parts are dry before reassembly.

To reassemble, center the Alignment Feature on the
Backpiece with the

Flow-Vu

® Indicator, as shown.

Press firmly to attach the Backpiece. For mouthpiece
models, the protective cap should always be placed
on the mouthpiece when the product is not in use.

5

CLEANING INSTRUCTIONS FOR MASK AND MOUTHPIECE CHAMBERS

THIS PRODUCT CAN BE USED RIGHT OUT OF THE PACKAGE AND THEN CLEANED WEEKLY.

© 2009 Monaghan Medical Corporation. ™ and ® are trademarks and registered trademarks used under license by Forest Pharmaceuticals, Inc. Printed in USA. Covered by one
or more of the following patents # 5,645,049; 5,848,588; 5,988,160; 6,293,279; 6,345,617; 6,435,177; 6,904,908 and patents pending. RMC 16416 Revision: 01/10

Subsidiary of Forest Laboratories, Inc.
St. Louis, MO 63045
Monaghan Medical Corporation, 5 Latour
Ave., Suite 1600, Plattsburgh, NY 12901

Distributed by:

Exhalation
Valve

Inhalation Valve

Cap

Backpiece

LARGE MASK CHAMBER

MOUTHPIECE CHAMBER

Flow-Vu

®

Inspiratory

Flow Indicator

Anti-Static Chamber

ComfortSeal

®

Mask

Alignment Feature

FlowSignal

®

Whistle

4

2

3

5

6

7

Apply mask to face and ensure an
effective seal. The

Flow-Vu

®

Inspiratory Flow Indicator only moves
if the patient has a good seal.

Breathe out gently and depress the MDI at the beginning of a slow inhalation. Use the

Flow-Vu

® Indicator to assist in the coordination of this maneuver. Maintain seal for 5 – 6

breaths after the MDI is depressed.

Slow down inhalation if you hear the

FlowSignal

®

Whistle sound.

Administer one (1) puff at a time.

1

2

3

4

Small Mask

Medium Mask

Mouthpiece

Large Mask

Notes:

• Storage and operating range 5° C - 40° C (41° F - 104° F) at

15 to 95% relative humidity.

• Product may need to be replaced after 12 months of use.

Environmental conditions, storage and proper cleaning can
affect product life span.

• THIS PRODUCT CONTAINS NO LATEX.
• Do not share this medical device.
• Clarity of the chamber is a result of the properties of the

StatBan

® anti-static material.

• If you notice medication build-up in your chamber, wash the

inside of the chamber gently with a soft cloth.

Cautions:

• To ensure proper performance this product should only

be cleaned according to these instructions.

• Do not leave the chamber unattended with children.
• Federal (USA) law restricts the sale of this device on or

by the order of a physician.