background image

19 

 

7.4

 

running the demonstrations programs 

Two demonstration programs can also be used to diagnose the device. The source code and an 
executable file for each of these demonstration programs are provided in two separate directories 
named 

\gravity

 and 

\torus

Once the system is setup, we suggest running application 

gravity

 to check that everything is 

working properly and to evaluate your system's performance independently of the graphics ren-
dering performance. Application 

torus

 will allow you to test the combined performance of hap-

tics and graphics rendering. 

gravity example 

This example program runs a best effort haptic loop to compensate for gravity. The appropriate 
forces are applied at any point in space to balance the device end-effector so that it is safe to let 
go of it. The refresh rate of the haptic loop is displayed in the console every second. 
 

 

figure 9 – gravity example 

Содержание omega haptic device Series

Страница 1: ...USER MANUAL omega x haptic device version 1 9 Force Dimension Switzerland www forcedimension com ...

Страница 2: ...2 ...

Страница 3: ...he Force Dimension SDK to describe the operation modes of the omega x haptic device glossary SDK refers to the Software Development Kit SDK for all Force Dimension products omega x refers to the base haptic device shared by the omega 3 omega 6 and omega 7 hap tic devices Unless specified all instructions in this manual apply to all three device types ...

Страница 4: ...4 ...

Страница 5: ... x under Linux 12 5 1 installing the software 12 5 2 installation description 12 5 3 installing the drivers 12 6 configuring the omega x under macOS 13 6 1 installing the software 13 6 2 installation description 13 6 3 installing the drivers 13 7 operating the omega x 14 7 1 coordinate system 14 7 2 operating modes 16 7 3 running the Haptic Desk program 18 7 4 running the demonstrations programs 1...

Страница 6: ...6 ...

Страница 7: ...a 3 haptic device 1 base plate 2 control unit 3 front arms 4 end effector 5 force button 6 calibration pole 7 calibration pit 8 status LED 9 force LED 10 user button 11 power switch 12 power connector 13 USB connector 2 3 1 5 9 8 2 1 11 12 13 10 6 4 7 ...

Страница 8: ...ructions marked on your haptic device 3 do not use or place your haptic device near water 4 place your haptic device securely on a stable surface 5 make sure that the workspace of your haptic device is free of objects 6 do not overload wall outlets and extension cords as this can result in a risk of fire or electri cal shock 7 switch off your haptic device when it is not in use 8 to reduce the ris...

Страница 9: ...SE KEEP THE ORIGINAL PACKAGING ONLY USE THE ORIGINAL PACKAGING DURING STORING OR SHIPPING 3 1 unpacking the device Before unpacking the omega x haptic device remove the haptic device foam stabilizer and the accessories foam section located inside the shipping box figure 2 view when opening the shipping box haptic device foam stabilizer haptic device accessories ...

Страница 10: ... USB cables as well as the USB flash drive figure 4 omega x accessories 3 2 installing the power supply Plug the power supply into the power connector For safety purposes you should only operate your omega x haptic device using the original Force Dimension power supply that came with your hap tic device controller Replacement power supplies can be ordered directly from Force Dimension end effector...

Страница 11: ...es required to run the omega x software The required DLL files are also copied to the Windows system folder during the installation drivers subfolder This directory contains the USB drivers required to operate your haptic device examples subfolder This directory contains the demonstration programs Example applications described in section 7 4 and come with their full source code doc subfolder All ...

Страница 12: ...s directory contains the demonstration executables and the binary files required to run the omega x software examples subfolder This directory contains the demonstration programs Example applications described in section 7 4 and come with their full source code doc subfolder All documentation files and notices are located in this subfolder manuals subfolder All hardware user manuals are located in...

Страница 13: ... following directories bin subfolder This directory contains the demonstration executables and the binary files required to run the omega x software examples subfolder This directory contains the demonstration programs Example applications described in section 7 4 and come with their full source code doc subfolder All documentation files and notices are located in this subfolder manuals subfolder ...

Страница 14: ...sed in Cartesian coordinate and in IUS metric unit Figure 5 illustrates the coordinate system The actual origin of the coordinate system 0 0 0 is located on a virtual point situated at the cen ter of the physical workspace of the haptic device figure 5 Cartesian coordinate system of the omega x haptic device X axis Y axis Z axis ...

Страница 15: ...nt sensors mounted of each revolute axis of the wrist figure 6 reference frame of the wrist omega 6 and omega 7 haptic devices gripper angle The angular position of the force gripper is returned in either degrees or radian A positive angle value is returned for right hand omega 7 haptic devices A negative angle value is returned for left hand haptic devices Angular values closer to zero correspond...

Страница 16: ...on pole in the dedicated calibration pit The device detects when the calibration position is reached and the status LED stops flashing Figure 7 illustrates the calibration procedure After the initial calibration described above the LED will stop flashing omega 3 figure 7 calibration procedure On the omega 6 and omega 7 the status LED will blink at a slower frequency indicating that the wrist is us...

Страница 17: ...every time the forces are disabled When the brakes are engaged a viscous force is created that prevents rapid movement of the end effector safety features The omega x features several safety features designed to prevent uncontrolled application of forces and possible damage to the device These safety features can be adjusted or disabled via a protected command in the Force Dimension SDK IMPORTANT ...

Страница 18: ...tic devices connected to the system test the position reading of the haptic device in Cartesian coordinates test all force and torque capabilities of the haptic device run the auto calibration procedure read the haptic device status read the haptic device encoder sensors individually read the haptic device user button if available figure 8 Haptic Desk test and diagnostic program ...

Страница 19: ...ing is working properly and to evaluate your system s performance independently of the graphics ren dering performance Application torus will allow you to test the combined performance of hap tics and graphics rendering gravity example This example program runs a best effort haptic loop to compensate for gravity The appropriate forces are applied at any point in space to balance the device end eff...

Страница 20: ...enGL scene with haptic feedback figure 10 torus example note OpenGL must be installed for your compiler and development environment to compile this example Please refer to your compiler documentation for more information or consult http www opengl org ...

Страница 21: ...p to 4 0 KHz power universal 100V 240V software platforms Microsoft Windows Linux all distributions Apple macOS Blackberry QNX WindRiver VxWorks libraries Haptics SDK Robotics SDK features ergonomics the device can be used with both left and right hands structure delta based parallel kinematics active gravity compensation calibration automatic driftless user input 1 user button safety velocity mon...

Страница 22: ...22 ...

Страница 23: ...r universal 100V 240V software platforms Microsoft Windows Linux all distributions Apple macOS Blackberry QNX WindRiver VxWorks libraries Haptics SDK Robotics SDK features ergonomics available in left and right hand configuration structure delta based parallel kinematics hand centered rotation movements decoupling between translation and rotation movements active gravity compensation calibration a...

Страница 24: ...24 ...

Страница 25: ...Hz power universal 100V 240V software platforms Microsoft Windows Linux all distributions Apple macOS Blackberry QNX WindRiver VxWorks libraries Haptics SDK Robotics SDK features ergonomics available in left and right hand configuration structure delta based parallel kinematics hand centered rotation movements decoupling between translation and rotation movements active gravity compensation calibr...

Страница 26: ...26 ...

Страница 27: ...27 ...

Страница 28: ...e information or product described herein This document may contain or reference information and products protected by copyrights or patents and does not convey any license under the patent rights of Force Dimension nor the rights of others All trademarks or trade names are properties of their respective owners Copyright 2020 Force Dimension All rights reserved ...

Отзывы: