Operating Instructions  

EMGZ310 

Compact digital measuring amplifier 

EMGZ310.ComACT 

Digital Tension Measuring Amplifier with Optional Operation 

via FMS-comACT App 

 

 

Document version 

2.32 

 

10/2020 NS 

 Firmware 

EMGZ310 

1.14 

 Firmware 

EMGZ310.ComACT  V2.10 

 

 

 

EMGZ310 EMGZ310.ComACT 

 

Diese Bedienungsanleitung ist auch in Deutsch erhältlich. 

Bitte kontaktieren Sie Ihre lokale FMS Vertretung. 

 

© by FMS Force Measuring Systems AG, CH-8154 Oberglatt – All rights reserved. 

 

 

Содержание EMGZ310

Страница 1: ...ptional Operation via FMS comACT App Document version 2 32 10 2020 NS Firmware EMGZ310 V 1 14 Firmware EMGZ310 ComACT V2 10 EMGZ310 EMGZ310 ComACT Diese Bedienungsanleitung ist auch in Deutsch erhältlich Bitte kontaktieren Sie Ihre lokale FMS Vertretung by FMS Force Measuring Systems AG CH 8154 Oberglatt All rights reserved ...

Страница 2: ...e of Delivery 7 4 QUICK GUIDE QUICK START 8 4 1 Preparations for Parameterization 8 4 2 Mounting Sequence 8 4 3 Installation and Electrical Connections 8 4 4 Force sensor installation 9 4 5 Electrical Connections 9 5 OPERATION AND CONFIGURATION 11 5 1 Power up the EMGZ 310 11 5 2 Operating panel and controls 11 6 CALIBRATION OF THE MEASURING SYSTEM 13 6 1 Offset Compensation 13 6 2 Calibration Adj...

Страница 3: ...erformance data can endanger the safety and health of persons Work with respect to operation maintenance retrofit repair or setting the device described here must only be performed by expert personnel 2 1 Presentation of Safety Information 2 1 1 Danger that Could Result in Minor or Moderate Injuries Danger warning caution Type of danger and its source Possible consequences of nonobservance Measure...

Страница 4: ...itive electronics Do not work with a blunt tool screw driver pliers etc on the housing Use suitable grounding grounding wrist strap etc when working on the electronics The devices should have a distance of at least 15 mm to one another in the control cabinet for proper cooling 2 3 Regulatory Notices NOTE This device complies with Part 15 of the FCC Rules and with Industry Canada licence exempt RSS...

Страница 5: ...accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense NOTE Changes or modifications made to this equipment not expressly approved by FMS Force Measuring Systems AG may void the FCC au...

Страница 6: ...es nuisibles lorsque l équipement est utilisé dans un environnement commercial Cet équipement produit utilise et peut émettre une énergie de fréquence radio et s il n est pas installé ni utilisé conformément au manuel d instruction il peut générer des interférences affectant les communications radio L utilisation de cet équipement dans une zone résidentielle peut causer des interférences nuisibles...

Страница 7: ...either three easily accessible keys or your smartphone tablet The free FMS ComACT app is available for download for IOS and Android devices and makes parameter setting and value reading as efficient and comfortable as never before 3 3 Scope of Delivery The following is included in the scope of delivery Measuring amplifier The following is not included in the scope of delivery AC DC power supply mi...

Страница 8: ...e obtained from the mounting instructions of the force sensors Connect the force sensors to the amplifier Connect the amplifier to the supply voltage The voltage supply must be in the range of 18 to 36 VDC Perform offset compensation and calibration Change the parameter settings as needed 4 3 Installation and Electrical Connections Warning To improve natural convection and keep heating of the ampl...

Страница 9: ...roducts The mounting instructions are included with the force sensors 4 5 Electrical Connections One or two force sensors can be connected to the amplifier When two sensors are used they are connected in parallel internally The force sensors and amplifier are connected using a 2 x 2 x 0 25 mm2 AWG 23 shielded twisted cable Figure 3 electrical connections EMGZ310_BA_Manual ai ...

Страница 10: ...locks use a small slotted screwdriver as a lever Warning Poor grounding can result in electric shocks for persons malfunctions of the overall system or damage to the measuring amplifier Proper grounding must always be ensured Note Cable shielding may only be connected to one side of the measuring amplifier On the side of the force sensor shielding must remain open ...

Страница 11: ...ion of the red point on the first sensor results in a positive output signal If not swap the two signal wires of this force sensor If applicable connect the second force sensor Check whether applying a force in measuring direction on the second sensor results in a positive output signal If not swap the two signal lines of this force sensor 5 2 Operating panel and controls Figure 5 front view EMGZ3...

Страница 12: ...ols Button Description Change the selections increase decrease the values or choose Scroll up or down the parameter lists Enter 0 PARA 3 Sec Start Offset compensation precedure CAL PARA 3 Sec Start calibration procedure PARA 3 Sec Parameter setting Table 1 controls ...

Страница 13: ...matically and stored in the amplifier 6 2 Calibration Adjusting the Gain Factor Calibration is used for matching the gain factor with the force sensors After calibration the displayed force corresponds the effective material tension of the web filament Two calibration methods are possible The first calibration method described here uses a defined weight There is also a calculation method for the g...

Страница 14: ...lue with Save the changes with Enter the weight GAIN connected to the rope Pay attention to the selected unit 10 kg would require an input of 100 N if the system is set to N You can modify the value with Save the changes with The parameter F mOut set the value of the indicated tension with maxium output 10 VDC or 20 mA on the display of the amplifier You can modify the value with Save the changes ...

Страница 15: ...tion EMGZ310 EMGZ310 ComACT 21 10 2020 15 Modify the value with until the voltage output correlates with the weight Save the changes with The display will return to the main screen The calibration procedure is complete ...

Страница 16: ...y of the filter The lower the cut off frequency the more sluggish will the output as well as the display be The filter stabilizes the output signal in case of highly fluctuating tension values A correctly tuned filter will straighten the displayed value Unit Hz Min 0 1 Max 999 9 Default 10 0 Unit This parameter determines the unit used in the system The label on the force sensor shows always the n...

Страница 17: ...and enter 12kN as system force NOTE If the system is set to voltage mode VoltGauge this parameter is disabled Unit N kN lb g kg selected in unit Min 1 Max 9999 Default 1000 F mOut This parameter determines the tension value N kN lb g kg that is indicated at the maximum output of the amplifier 10 V or 20mA NOTE If lb pound is selected the system switches from metric to imperial measuring units Sele...

Страница 18: ...10 ComACT 21 10 2020 18 Mode The Device Mode determines the functionality of the amplifier If set to Volt Gauge force values will be displayed only in Volt V or mA Force units will not be available anymore Selection kgkNN V Default kgkNN ...

Страница 19: ...ild jpg Only a single mobile device at a time can connect to an amplifier As soon as the connection between amp and mobile device is setup the display of the amp will start flashing The FMS ComACT app is meant as an operation aid to display actual readings and to help configure the amplifier The settings are only stored in the amplifier and not in the mobile device 8 1 Configuration via the App Th...

Страница 20: ...libration leads you to the offset and gain menu Button for the export of the parameter settings You can perform an offset without any further input Make sure that there is no web applied to the roller Successful input and operations are confirmed Before you perform the calibration you have to enter the value for the corresponding force that you have applied to the measuring roller Table 2 Screensh...

Страница 21: ...asing amplifier at all times Start recording Selection of the interval and resolution measuring values per second Recording in progress with remaining time and actual reading You can determine the recording at all time Recorded data will remain available for the export End of a measurement Measurement time expired Button Export for further use of the measurement data Transmission of measurement da...

Страница 22: ...Z310 ComACT 21 10 2020 22 The name of the export csv file contains the following information Figure 8 file name Note In each export file you will find the parameter settings of the amplifier This can be of help to document the settings ...

Страница 23: ...Operating Instruction EMGZ310 EMGZ310 ComACT 21 10 2020 23 9 Dimensions Figure 9 dimensions ...

Страница 24: ...IP 20 Power supply 24 VDC 18 to 36 VDC Power consumption 5 W Temperature range 10 to 50 C 14 to 122 F Weight 370 g 0 82 lbs FMS Force Measuring Systems AG Aspstrasse 6 8154 Oberglatt Switzerland Tel 0041 1 852 80 80 Fax 0041 1 850 60 06 info fms technology com www fms technology com FMS USA Inc 2155 Stonington Avenue Suite 119 Hoffman Estates IL 60169 USA Tel 1 847 519 4400 Fax 1 847 519 4401 fmsu...