FMC INVALCO WCM 7300 Скачать руководство пользователя страница 1

Issue/Rev. 0.2 (7/00)

Page 1

Bulletin MNIS005

Model WCM 7300

October 2000

INVALCO

Installation/Operation Manual

Water Cut Monitor with

Temperature Compensation

Table of Contents

General Overview..............................................................................................................................2-3

Specifications and Definitions...............................................................................................................................3

Error Messages...............................................................................................................................................4

Product Temperature Compensation.........................................................................................................................4

Set Point and Loop Calibration......................................................................................................................5

Program Flow Chart..........................................................................................................................................5

Quick Steps....................................................................................................................................................6

Programming of WCM 7300 ....................................................................................................................7

Installation..................................................................................................................................7

Electrical Connections.......................................................................................................................................8

Trouble Shooting............................................................................................................................................9-10

Dimensions..................................................................................................................... ..........11

FMC INVALCO 

 Fluid Control 

 P.O. Box 1377

 

 Stephenville, Texas 76401 

 Telephone: 254/968-2181 

 FAX 254/968-5709 

 

Toll Free: 800/468-2526

Содержание INVALCO WCM 7300

Страница 1: ...iew 2 3 Specifications and Definitions 3 Error Messages 4 Product Temperature Compensation 4 Set Point and Loop Calibration 5 Program Flow Chart 5 Quick Steps 6 Programming of WCM 7300 7 Installation 7 Electrical Connections 8 Trouble Shooting 9 10 Dimensions 11 FMC INVALCO Fluid Control P O Box 1377 Stephenville Texas 76401 Telephone 254 968 2181 FAX 254 968 5709 Toll Free 800 468 2526 ...

Страница 2: ...tween 150 and 275 depending on the gravity of the oil residual production drilling chemicals wa ter cut gas content etc This reading can be valuable if you ever need to trouble shoot your system It should be recorded at initial installation for future reference It should be a steady reading A change of 4 to 5 Pf will change the read ing of the 7300 by 1 If anything varies in th process stream maki...

Страница 3: ...ill toggle the flashing cursor between N and Y Selecting Y and then Enter commits you to making a grindout or cut entry N will put you into the Manual TC Entry Mode Cal With X Cut Pressing the up arrow allows selection of either one or two cuts Select 1 for one cut calibra tion Select 2 for two cut calibration with auto TC Enter Cut X XX Enter cut or grindout value for cali bration Manual TC XXXX ...

Страница 4: ...il condition red LED flashing and release of the relay to bypass solid red LED The reverse condition green LED and transition to good oil will occur in 1 10th the delay time Undo Changes YN This allows the opportunity to reverse or undo all calibration changes made during the past programming session The selection is auto matically N unless Y is deliberately selected Y should be used after an Auto...

Страница 5: ...s Below is a flow chart programming guide Each bubble or diamond represents what the display will show The information adjacent to the small circles between the bubbles instruct you on how to proceed through the programming modes Press All Three Buttons Program Mode Press Left arrow 3 times Up arrow 2 times 20 mA X XX Set pt X XX Delay XXX Undo Changes Program Done Cut X XX Last Cut X XX Do Cal Ca...

Страница 6: ...eading something different To start over press Enter until past the display that says Program Done Then restart at Step 3 7 Press Enter to proceed to the next display 8 The display will now say Do Cal YN The N should be flashing meaning no Select Y yes by pressing the Left Arrow key The Y will now be flashing 9 Press Enter to confirm yes and proceed to the next display 10 The display should now re...

Страница 7: ... value becomes a bad oil reading The programmed delay sets the delay in seconds between a bad oil condition and relay release The indicator LED is green when the cut is below the setpoint and red when at or above the setpoint When the LED flashes red the setpoint is exceeded but the delay time has not expired When the relay goes to bypass delay is over the LED will go to solid red The same applies...

Страница 8: ...ct the terminal of the power supply to the Loop Terminal To connect the 4 20 mA analog output as non isolated Figure 7 Install a jumper wire from 24V to Loop 24 24 GND 5V SIG SIG GND LOOP LOOP NO NC C 4 20 mA Connect a jumper wire from the 24 Terminal to the Loop Terminal Connect the Loop Terminal to the terminal of the monitoring equipment Connect the terminal of the monitoring equipment to the 2...

Страница 9: ...e dump valve when moni toring oil from a pressure vessel as solution gas will break out as the fluid goes from the higher vessel pressure to the lower pressure downstream of the valve An increase in gas will make the reading become smaller Also since the size and volume of the gas bubbles is constantly changing with flow the probe value can be very un stable as it attempts to follow the fluc tuati...

Страница 10: ...is too low for calibration and that the instrument will go into Under Range if the value goes much lower If the Dk should continue to drop causing the probe value to decrease even further the Under Range message will come on The LED will blink red for the time delay setting and then go solid red The relay will actuate to the oil by passing position Note that the probe value numbers that you can br...

Страница 11: ...Probe Dimensions Inches mm No Size A B C D E F G 81017222 2 150 17 0 432 8 5 216 4 5 114 4 5 114 8 0 203 5 3 135 8 0 203 81017232 3 150 32 0 813 16 0 406 4 5 114 4 5 114 8 0 203 5 3 135 8 3 211 81017242 4 150 32 0 813 16 0 406 4 5 114 4 5 114 8 0 203 5 3 135 8 8 224 81017262 6 150 32 0 813 16 0 406 4 5 114 4 5 114 8 0 203 5 3 135 9 9 251 ...

Страница 12: ... the specifications are currently in effect Otherwise the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect FMC INVALCO Fluid Control P O Box 1377 Stephenville TX 76401 Phone 254 968 2181 FAX 254 968 5709 Toll Free 800 468 2526 Printed in U S A 10 00 FMC INVALCO All rights reserved Issue Rev 0 2 10 00 ...

Отзывы: