12

TempRange

When measuring temperature, you need to use the appropriate temperature range, which 

depends on the target temperature. The Meter has these temperature ranges: 

-20 °C  to  150 °C  (-4 °F  to  302 °F)

100 °C to 400 °C (212 °F to 752 °F)

The Meter can automatically use the appropriate range based on the target temperature if you 

set the TempRange to 

Auto

 in the Main Menu. 

Emissivity

All objects radiate infrared energy. The actual surface temperature and emissivity of the target 

affect the quantity of energy radiated. The Meter senses the infrared energy from the surface 

of the target and uses the data to calculate an estimated temperature value. Many common 

materials such as wood, water, skin, cloth, and painted surfaces, including metal, radiate 

energy well and have a high emissivity factor of 

90 % (or 0.90). The Meter measures 

temperatures accurately on targets with a high emissivity.
Shiny surfaces or unpainted metals do not radiate energy well and have a low emissivity factor 

of 

<

0.60. For the Meter to calculate a more accurate estimate of the actual temperature of 

targets with a low emissivity, adjust the emissivity setting.

W

 

Warning

To prevent personal injury, see emissivity information for actual temperatures. 

Reflective objects result in lower than actual temperature measurements. These 

objects pose a burn hazard.

Set emissivity as a direct value or from a list of emissivity values for some common materials.

Note

Surfaces with an emissivity of 

<

0.60 make it difficult to determine reliable and 

consistent actual temperatures. The lower the emissivity, the greater the potential of 

error when the Meter calculates the temperature measurement because more of the 

energy reaching the camera is specified as background temperature. This is also true 

when adjustments to the emissivity and reflected background adjustments are 

performed properly.

To adjust emissivity: 
1. In the Emissivity submenu, push 

W

 or 

X

 to select which emissivity to use, or select 

SelfDefine

2. If you select 

SelfDefine

, push 

O

 and 

W

 or 

X

 to manually set the emissivity. 

3. Push 

O

 again to save.

4. Push 

P

 to return to the previous menu.

Содержание VT06

Страница 1: ...sual IR Thermometer Users Manual October 2021 English 2021 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respectiv...

Страница 2: ...warranty service contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty post...

Страница 3: ...ration 7 Power On and Power Off the Product 7 Focus 7 Laser Pointer 7 Camera mode 7 Capture and Save an Image 8 Save Images to a PC 8 Main Menu 9 Gallery 11 Review 11 Delete all images 11 TempRange 12...

Страница 4: ...ii VT06 VT08 Users Manual...

Страница 5: ...fer images and data stored in internal memory to a PC through a direct USB connection to the PC A rugged rechargeable lithium ion smart battery provides power to the Meter Direct ac power is accessibl...

Страница 6: ...different features some of the information in the manual might not apply to your Meter Note Unless otherwise specified the figures in this document show the VT08 Visual IR Thermometer Features Table 1...

Страница 7: ...Visual IR Thermometer Product Familiarization 3 Table 3 shows the back of the Products Table 2 Front of VT08 Item Description A Display screen B See Control Panel C USB Cable Power input 1 3 2...

Страница 8: ...roduct For the VT08 device only you can also use the O button to turn on or turn off the LED Light Table 4 shows the functions of the buttons on the Control Panel Table 3 Back of the Products VT08 VT0...

Страница 9: ...t VT08 only Within a menu or submenu Push to save the change and go back to the previous menu Push to perform the function P Push and hold for 2 seconds to power on or power off the Product Push to ca...

Страница 10: ...Item Description A Center point temperature G Temperature range higher value B Region of Interest Measure ROIMeasure area This area outlines the high and low temperature markers in its spot box H ROI...

Страница 11: ...between environments with large differences in ambient temperature allow for additional adjustment time Note If the Product crashes push and hold P for 3 seconds to power off the Product Focus The VT...

Страница 12: ...nds pull and release the Trigger again or push O to save the image If you decide to discard the image wait longer than 2 seconds to push O or push P Save Images to a PC To save images to a PC with the...

Страница 13: ...Trigger to return to Camera mode Table 6 Main Menu Options Menu Option Function Gallery View or delete an image Use Review to look at an image in the internal memory storage You can delete this image...

Страница 14: ...t the measure area to help with temperature measurement See ROIMeasure Alarm Set the alarm parameters See Alarm Reflmeasure Set the reflected temperature Distance The farther the distance the weaker t...

Страница 15: ...lery submenu push W or X to highlight Delete all images 2 Push O and push W or X to highlight then push O again to delete all images DateTime Set the date and time See DateTime Language Reset Reset th...

Страница 16: ...l and have a low emissivity factor of 0 60 For the Meter to calculate a more accurate estimate of the actual temperature of targets with a low emissivity adjust the emissivity setting W Warning To pre...

Страница 17: ...bmenu push O to turn on or turn off the alarm 2 When the alarm is on a Push X to highlight High b Push O to turn on the high temperature alarm c Push X to highlight High temperature value d Push O to...

Страница 18: ...h W or X to find the year you want 3 Push O to save the setting 4 To set the month and day follow the first step 1 through step 3 in this section using step 1 to highlight Month and Day as appropriate...

Страница 19: ...nsparency Levels Fluke VT Report Desktop Software Fluke VT Report desktop software for a PC is available to use with the Product and contains features to analyze images organize data and information a...

Страница 20: ...ective coating Do not clean the lens too vigorously because this can damage the anti reflective coating To clean the lens 1 Use a pressurized can of air or a dry nitrogen ion gun if available to blow...

Страница 21: ...re you connect it to the charger See the charging temperature specification Do not charge in hot or cold areas When you charge in extreme temperatures battery capacity might be decreased When you char...

Страница 22: ...VT06 VT08 Users Manual 18...

Отзывы: