3 Technical data
3.1 System module - VIB 7.900
Parameter
VIB 7.900
INPUTS AND OUTPUTS
Analog inputs
6 synchronous channels
Type of signal, channel
distribution
6 x U / IEPE (ICP), selectable
Type of sensor
IEPE sensor; Sensor with voltage output
Tacho / pulse input
Frequency input: ±30V; Threshold ±30V DC (default 2.5V)
Tacho / pulse output
Sensor voltage supply
Digital inputs
Optocoupler input: -3V to +30V, switching thresholds 6.5V to 8.5V DIN EN 61131-2
operating range DC 24V type I and II
Digital outputs
Relay turnkey: 30V DC / 30V AC / 2A DIN EN 61131-2 utilization category AC15 and
DC13
System OK output
Relay opener: 30V DC / 30V AC / 2A DIN EN 61131-2 utilization category AC15 and
DC13
Ethernet
Data rate: 100 MBit, half duplex
Serial interface
RS232, 115200 baud
Services
Modbus-TCP
LED indicators
6x analog IN, 1x tachometer IN, 1x system, 1x event, 2x Ethernet, 1x digital IN, 1x sys-
tem OK, 1x digital OUT
MEASUREMENT
Dynamic range
110 dB @ 24 bit
Sampling rate
131 kHz / 50 kHz range
FFT lines
6400 (default), 102400 (analysis)
Measuring range
± 24V
GENERAL
Ambient temperature
Operation: -20°C ... +70°C [ -4 °F ...+ 158 °F]
Storage: -40°C ... +80°C [ -40 °F ...+ 176 °F]
Relative humidity
max. 95 % (25°C [77°F], no condensation)
System supply
24±6 VDC / 0.25 A
Memory
Flash: 2 GB, RAM: 256 MB
Housing material
Aluminum
VIBGUARD compact
13
Installation and Operation
Содержание VIBGUARD compact
Страница 6: ...Empty page...
Страница 49: ......