RUSKA 7750i
Users Manual
6-8
The screen will display the current status of the pressure sensor Stability and
Temperature.
If any of the above are unstable, then the system will delay until stability is achieved.
Pressing OK [F6] will bypass this wait period.
Note
Bypassing this wait period can have a negative effect on the zeroing
procedure.
Qc Channel
1.
Verify that the Ps and Qc sensors are close to atmospheric pressure.
2.
Enter the Calibration screen by selecting MENU | MENU | CALIBRATION.
3.
Select Zero [F1]. Do NOT press the Calibrate button.
gme30.bmp
Figure 6-3. Menu | Menu | Calibration
4.
Do not disturb the instrument while zeroing is in process.
5.
Wait for the zeroing procedure to finish.
gme31.bmp
Figure 6-4. Zeroing Screen
6.
Press PREVIOUS to return to the Main Menu.
Содержание RUSKA 7750i