Ps PR IMAR Y
S E NS OR
Qc PR IMAR Y
S E NS OR
CONTR OL
S E NS OR
CONTR OL
S E NS OR
PR E S S UR E
R E GULATOR
VACUUM
S UPPLY
PR E S S UR E
S UPPLY
Ps TE S T
POR T
Pt TE S T
POR T
VE NT
Ps APPLY
Pt E QUALIZE
VE NT
Pt TE S T
Ps TE S T
Pt APPLY
Ps R E LE AS E
Pt R E LE AS E
ZE R O
gme04.eps
Figure 2-3. RUSKA 7750i ADTS - Pneumatics Schematic Diagram
Go to Ground or Vent Procedure
The Go-To-Ground is implemented as a multi-step procedure:
a.
Control Ps to close to Atmosphere and Qc to zero at the current rate.
b.
Measure actual atmospheric pressure.
c.
Control Ps to measured atmosphere.
d.
Open Ps Vent and Pt Equalize valves.