RUSKA 7252
Users Manual
4-22
2.
The program is entered from the
Program Menu
. From the
Main Menu
(press
PREVIOUS
until the Main Menu appears), press
Menu
[
F6
]
| Program
[
F4
].
glb30.bmp
Figure 4-18. Channel A: Menu | Program - Menu
3.
Use the rotary knob to highlight “new”.
4.
Press
Edit
[
F4
]. Since “new,” was highlighted, the Calibrator will create a new
program and give a default name of NAMEnn, where nn is a two digit number. The
program editing screen will appear, displaying the first step.
glb31.bmp
Figure 4-19. Channel A: Menu | Program | Edit - Menu
5.
Using the numeric keypad, enter the values for
Pressure
,
Tolerance
,
Dwell time
,
and
Max time
, pressing
ENTER
after each value. The rotary knob may be used to
skip fields.