Short Form Specification
Test
Sensitivity and Frequency Range
– Recall the DEFAULT settings.
– Press
INPUT A
.
– Select
50
W
input impedance,
1x
attenuation,
MAN
ual
trigger and
Trig
ger level
0 V
.
– Connect a signal from a HF generator to a BNC power
splitter.
– Connect the power splitter to Input A of your counter
and an oscilloscope.
– Set the input impedance to 50
W
on the oscilloscope.
– Adjust the amplitude according to the following table.
Read the level on the oscilloscope. The timer/counter
should display the correct frequency.
– Connect the signal to Input B.
– Press
INPUT B
.
– Select
50
W
input impedance,
1x
attenuation,
MAN
ual
trigger and
Trig
ger level
0 V
.
– Press
MEAS FUNC
®
Freq
®
Freq A
®
B
– Repeat the measurements above for Input B.
Voltage
– Recall the DEFAULT settings.
– Press
MEAS FUNC
®
Volt
®
Vpp
®
A
– Press
INPUT A
and select
DC
coupling. Do not apply an
input signal to Input A yet.
– Press
EXIT/OK
.
– The display should now indicate
(disregard the main parameter V
PP
):
V
MIN
= 0 ± 0.015 V and
V
MAX
= 0 ± 0.015 V
– Adjust the current limit of the voltage source to
<200 mA.
– C2.500 V
DC
to Channel A, using the external
low-pass filter on the input.
– The display should now indicate:
V
MIN
= 2.500 ± 0.040 V and
V
MAX
= 2.500 ± 0.040 V
– Repeat the measurement with inverted polarity.
– Press
INPUT A
and select
10x
.
– Press
EXIT/OK
.
CAUTION: Before the next step, make sure
the input impedance is still 1 M
W
. Apply-
ing more than 12 V without proper current
limiting may cause extensive damage to
the main PCB, if the impedance is set to
50
W
.
– Change the DC level to +50.00 V.
– The display should now indicate:
V
MIN
= 50.00 ± 0.65 V and
V
MAX
= 50.00 ± 0.65 V
– Repeat the measurement with inverted polarity.
– Disconnect the DC voltage from Channel A.
– Remove the external low-pass filter.
– Press
INPUT A
and select
1x
.
– Press
EXIT/OK
.
– Connect a sinusoidal signal to Input A with an amplitude
of 4.000 Vpp and a frequency of 100 kHz.
– The display should now indicate: 4.000 ± 0.150 V
PP
.
– Press
INPUT A
and select
10x
.
– Press
EXIT/OK
.
– Change the amplitude to 18.00 V
PP
.
– The display should now indicate:
18.00 ± 0.84 V
PP
.
– Disconnect the signal from Channel A.
– Press
MEAS FUNC
®
Volt
®
Vpp
®
B
– Press
INPUT B
and select
DC
coupling. Do not apply an
input signal to Input B yet.
– Press
EXIT/OK
.
– The display should now indicate
(disregard the main parameter V
PP
):
V
MIN
= 0 ± 0.015 V and
V
MAX
= 0 ± 0.015 V
– Proceed by repeating the measurements for Input B as
described above for Input A.
Trigger Indicators vs. Trigger Levels
NOTE: This test must be performed in the sequence given.
– Recall the DEFAULT settings.
2-4
Performance Check, Short Form Specification Test
Trigger Level
(manually set)
Trigger Indicator
Pass
Input A
Input B
+1 V
off
-1 V
on
0.0 V
blinking
Table 2-4
Trigger indicator check.
Frequency
(MHz)
Level
Pass/Fail
mV
rms
dBm
Input A
Input B
10
15
–23
50
15
–23
100
15
–23
200
15
–23
300
25
–19
Table 2-5
Sensitivity for inputs A & B at various frequencies
Содержание PM6690
Страница 1: ...Timer Counter Analyzer PM6690 Service Manual...
Страница 2: ...II 4822 872 20306 May 2006 Fourth Edition Fluke Corporation All rights reserved Printed in Sweden...
Страница 4: ...This page is intentionally left blank IV...
Страница 7: ...Chapter 1 Safety Instructions...
Страница 11: ...Chapter 2 Performance Check...
Страница 18: ...This page is intentionally left blank 2 8 Performance Check RF Options...
Страница 19: ...Chapter 3 Disassembly...
Страница 23: ...Chapter 4 Circuit Descriptions...
Страница 24: ...This page is intentionally left blank 4 2...
Страница 41: ...Chapter 5 Corrective Preventive Maintenance...
Страница 44: ...This page is intentionally left blank 5 4 Other Important Information...
Страница 45: ...Chapter 6 Repair...
Страница 46: ...This page is intentionally left blank 6 2...
Страница 104: ...This page is intentionally left blank 6 60 Safety Inspection and Test After Repair...
Страница 105: ...Chapter 7 Calibration Adjustments...
Страница 119: ...Chapter 8 Replacement Parts...
Страница 146: ...This page is intentionally left blank 8 28 Replacement Parts Parts Common to Version A and Version B...
Страница 147: ...Chapter 9 Schematic Diagrams...
Страница 150: ...This page is intentionally left blank 9 4 Schematic Diagrams...
Страница 151: ...This page is intentionally left blank Schematic Diagrams 9 5...
Страница 152: ...Main Board PCB 1 Component Layout 9 6 Schematic Diagrams...
Страница 154: ...9 8 Schematic Diagrams This page is intentionally left blank...
Страница 156: ...9 10 Schematic Diagrams This page is intentionally left blank...
Страница 158: ...9 12 Schematic Diagrams This page is intentionally left blank...
Страница 160: ...9 14 Schematic Diagrams This page is intentionally left blank...
Страница 162: ...9 16 Schematic Diagrams This page is intentionally left blank...
Страница 164: ...9 18 Schematic Diagrams This page is intentionally left blank...
Страница 166: ...This page is intentionally left blank 9 20 Schematic Diagrams...
Страница 168: ...Display Keyboard PCB 2 Component Layout 9 22 Schematic Diagrams...
Страница 170: ...This page is intentionally left blank 9 24 Schematic Diagrams...
Страница 172: ...This page is intentionally left blank 9 26 Schematic Diagrams...
Страница 173: ...Main Board PCB 1 Component Layout Bottom Side Schematic Diagrams 9 27...
Страница 174: ...Main Board PCB 1 Component Layout Top Side 9 28 Schematic Diagrams...
Страница 176: ...9 30 Schematic Diagrams This page is intentionally left blank...
Страница 178: ...9 32 Schematic Diagrams This page is intentionally left blank...
Страница 180: ...9 34 Schematic Diagrams This page is intentionally left blank...
Страница 182: ...9 36 Schematic Diagrams This page is intentionally left blank...
Страница 184: ...9 38 Schematic Diagrams This page is intentionally left blank...
Страница 186: ...9 40 Schematic Diagrams This page is intentionally left blank...
Страница 188: ...9 42 Schematic Diagrams This page is intentionally left blank...
Страница 190: ...Display Keyboard PCB 2 Component Layout 9 44 Schematic Diagrams...
Страница 192: ...9 46 Schematic Diagrams This page is intentionally left blank...
Страница 193: ...Chapter 10 Appendix...
Страница 196: ...This page is intentionally left blank 10 4 Glossary...
Страница 197: ...Chapter 11 Index...
Страница 200: ...This page is intentionally left blank 11 4 Index...