Page 91
© 2011 Fluke Calibration
The display identifies the active piston-cylinder module. Press
[ENT]
to view the
active piston-cylinder module or press
[P-C]
or
[
←
]
or
[
→
]
to cycle through the
other piston-cylinder modules available (see Section 3.10.2). When the desired
piston cylinder module is identified, press
[ENT]
to proceed with viewing piston-
cylinder module information. Successive pressing of
[ENT]
steps through the
piston-cylinder information screens in the same order as the add and edit
functions (see Section 3.12.1, View Piston-Cylinder Module, Principle
3.12.1.4 Delete a Piston-Cylinder Module
).
PURPOSE
To delete a piston-cylinder module file so that it is no longer included in the
piston-cylinder modules available for selection by pressing
[P-C]
.
OPERATION
Caution
Once a piston-cylinder module file has been deleted it cannot
be recovered. To recreate it, the add piston-cylinder module
function must be used (see Section 3.12.1.1) which requires
entering all the piston-cylinder module information. Before
deleting a piston-cylinder module, consider editing it (see
Section 3.12.1.2).
To delete a piston-cylinder module press
[SPECIAL]
,
<1PC/MS>
and select
<1PC>
,
<4delete>.
The display identifies the active piston-cylinder module. Press
[P-C]
or press
[
←
]
or
[
→
]
to advance through the piston-cylinder modules available (see Section
3.10.2). When the desired piston cylinder module is identified, press
[ENT]
to
proceed with deleting the piston cylinder module information. Confirmation to delete
is required. Select
<2yes>
to delete the piston-cylinder module. Select
<1no>
to
NOT delete.
Note
The active piston-cylinder module (piston-cylinder module that is
currently selected) cannot be deleted. Before attempting to delete
a piston-cylinder module, be sure it is not currently selected.
3.12.1.5 Select The Active Piston-Cylinder Module
PURPOSE
To select the piston-cylinder module that is to active and to be used by PG9000
in calculating the defined pressure.
OPERATION
To select the active piston-cylinder module press
[SPECIAL]
,
<1PC/MS>
and
select
<1PC>
,
<5select>
. The display identifies the active piston-cylinder
module. Press
[P-C]
or press
[
←
]
or
[
→
]
to advance through the piston-cylinder
modules available (see Section 3.10.2). When the desired piston cylinder module
is identified, press
[ENT]
to make active.
3.12.1.6 Add a Mass Set
PURPOSE
To create a new mass set that will be available for selection as an active mass
set. Use this function when a new mass set has been acquired or a mass set
was deleted and must be reentered.
Содержание PG9000 Series
Страница 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Страница 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Страница 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Страница 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Страница 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...