Страница 3: ...pressed Liquid 2 Personal Protective Equipment 2 Symbols Used in this Manual 2 Connections 3 Connect the Gauge 4 Connect the Test Port Insert 4 Operation 5 Priming 5 Operation 6 Maintenance and Servicing 6 Screw Press Seals 6 Complete Disassembly 7 Ancillary Equipment 11 Angle Adapter P5543 11 Pointer Remover Punch P5551 11 ...

Страница 4: ...P5514 P5514 V P5514 EP Users Manual ii ...

Страница 5: ...iii List of Tables Table Title Page 1 Operating Fluid Compatibility 2 2 Symbols 3 3 Test Port Insert Parts List 5 4 Seal Kit Part Numbers 6 5 Screw Press Assembly Parts List 10 ...

Страница 6: ...P5514 P5514 V P5514 EP Users Manual iv ...

Страница 7: ...v List of Figures Figure Title Page 1 Connecting the Gauge 4 2 Test Port Insert 5 3 Reservoir and Test Port Assemblies 9 4 Screw Press Assembly 9 5 Angle Adapter 11 6 Pointer Remover Punch 11 ...

Страница 8: ...P5514 P5514 V P5514 EP Users Manual vi ...

Страница 9: ...is document refer to any pressure measuring instrument such as Transfer Standards Digital Calibrators and Transducers This system is only as accurate as the Master Test Gauge used The Master Test Gauge must be regularly calibrated on a Primary Reference Standard such as a Deadweight Tester to ensure accuracy is maintained The comparison test pump should be mounted securely to a stable workbench or...

Страница 10: ...ons that are not related to any specific procedures and do not appear elsewhere in this publication These are recommended precautions that personnel must understand and apply during equipment operation and maintenance to ensure safety and health and protection of property Compressed Liquid Use of compressed liquids can create an environment of propelled foreign matter Pressure system safety precau...

Страница 11: ...r fluids and protect the system from particulates we recommend the use of a Liquid to Liquid Separator see Section 5 Ancillary Equipment W Warning DO NOT use Teflon PTFE tape on these connections as this will prevent correct sealing The Gauge Adapter sealing system is designed for hand tight sealing up to 20 000 psi 1 400 bar wrenches or similar tools are not required over tightening can cause dam...

Страница 12: ... the position to face forward hold the gauge adapter and turn the instrument COUNTER CLOCKWISE so that it faces forward 4 Hold the instrument steady while turning the gauge adapter COUNTER CLOCKWISE until it pulls down onto the O ring Connect the Test Port Insert For devices with 1 8 BSP or NPT mounting threads the diameter of the thread is very close to the effective sealing diameter of the O rin...

Страница 13: ... Open reservoir valve 13 one turn counter clockwise and turn screw press fully in 2 Close valve 13 and turn screw press fully out 3 Open valve and turn screw press fully in Note During this operation bubbles may appear in the reservoir as trapped air is expelled For large volumes repeat steps 2 and 3 until no further bubbles appear 4 With valve open turn screw press fully out and close valve The t...

Страница 14: ...aution Never release the system pressure by using the reservoir valve alone as sudden depressurization will shock the system which may cause damage to sensitive instruments attached to it Maintenance and Servicing The Figures on the following pages detail the components of each assembly together with the relevant part numbers Where ASSY appears as a part number this indicates that this particular ...

Страница 15: ...ot in the barrel 11 Re tighten the barrel union nut W Caution If the lead screw assembly 24 shows signs of excessive wear then it is very likely that the associated components have worn also therefore the screw press assembly is available as a spare part Complete Disassembly 1 Remove the screw press assembly as described above 2 Unscrew and remove valve stem 13 taking care not to lose spring 12 an...

Страница 16: ...in the barrel is worn also If the bore is worn or scratched it will not seal correctly and may leak under pressure 11 Re assembly is the reverse of the above Ensure that all seals and sealing surfaces are clean and undamaged 1 1 3 3 4 4 5 6 7 8 9 10 11 12 13 14 14 15 2 2 gjq003 eps Figure 3 Reservoir and Test Port Assemblies ...

Страница 17: ...Hydraulic Comparison Test Pump Maintenance and Servicing 9 20 22 23 24 25 25 26 26 21 16 17 19 18 gjq004 eps Figure 4 Screw Press Assembly ...

Страница 18: ...alve Body 3923785 7 Bonded Seal 3862349 3922096 3922043 8 Reservoir 3919463 9 Locknut 3918244 10 Reservoir Cover 3918350 11 Washer 3916458 12 Spring 3922786 13 Valve Stem Assembly 3885982 14 Screw 3909422 15 Pump Stand 3918271 16 Manifold 3918280 17 O Ring 3864711 3865359 3865576 18 Locknut 3917862 19 Barrel 3919630 20 O Ring 3864782 3865214 3865541 21 Back Up Ring 3920516 22 Rambler 3919611 23 Ba...

Страница 19: ...hould be used The angle adapter fits directly onto the test station converting it through 90 degrees allowing the standard adapters to be used gjn022 bmp Figure 5 Angle Adapter Pointer Remover Punch P5551 To remove and refit the pointer of a pressure gauge use the pointer remover punch instrument This tool has a spring loaded plunger to quickly and consistently refit the pointer gjn023 bmp Figure ...