2640A/2645A Environmental Specifications
1-12.
Table 1-6 provides a summary of the environmental specifications for the 2640A/2645A.
Table 1-6. Environmental Specifications
Specification
Characteristic
Warmup Time
1 hour to rated specifications -or- 15 minutes if relative humidity
(noncondensing) is 50% or less.
Operating Temperature
-10°C to 60°C (14°F to 140°F)
Storage Temperature
-40°C to +70°C (-40°F to +158F)
Relative Humidity
90% maximum for -10°C to 28°C (14°F to 82.4°F)
75% maximum for 28°C to 35°C (82.4°F to 95°F)
50% maximum for 35°C to 60°C (95°F to 140°F)
(3 M
Ω
range, reduce humidity rating by 25% for 1 hour warmup.
The 3 M
Ω
range meets full humidity ratings with 2-hour warmup.)
Altitude
Operating: 2,000m (6,561 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum
Vibration
0.7g at 15 Hz
1.3g at 25 Hz
3g at 55 Hz
Shock
30g half-sine per Mil-T-28800
Bench handling per Mil-T-28800
2640A/2645A Input/Output Capabilities
1-13.
The following specifications include the input/output functions, including the Digital
I/O, Trigger Out, Trigger In, and Master Alarm output.
Digital I/O
1-14.
Table 1-7 provides a summary of the Digital I/O specifications for the 8 Digital I/O lines
(0 to 7). Digital I/O is located on the DIGITAL I/O connector, terminals 0 to 7, and
GND.
Table 1-7. 2640A/2645A DIGITAL I/O Specification
Specification
Characteristic
Maximum Input Voltage
30V
Minimum Input Voltage
-4V
Isolation
None (dc coupled)
Threshold
1.4V
Hysteresis
500 mV
Specification
Characteristic
Output Voltage - TTL Logical Zero
0.8V maximum for an Iout of -1.0 mA (1 LSTTL load)
Output Voltage - TTL Logical One
3.8V minimum for an Iout of 0.05 mA (1 LSTTL load)
Output Voltage - Non-TTL Load Zero
1.8V maximum for an Iout of -20 mA
Output Voltage - Non-TTL Load One
3.25V maximum for an Iout of -50 mA
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...