NetDAQ
Service Manual
5-24
Table 5-10. Relating Selftest Errors to Instrument Problems (cont)
Error
Code
Error Code
Description
Suspect
Assembly
Error Code
Discussion
12
Inguard open
thermocouple
detect failed.
A3 A/D
Converter
PCA
Background The open thermocouple detect circuit
checks on the amount of a 19.2 kHz voltage developed
across a “thermocouple.” Similar to the overload
detection, the self-test does not set any channel relays
and the detection circuit should detect the “open”
thermocouple.
Failure For this error to occur, the open-thermocouple
circuit has failed to detect the simulated open
thermocouple condition.
Correction Check the circuitry formed by the open
thermocouple detect circuit formed by A3U32 peak
detector and comparator and associated components
operates by applying a 19.2 kHz clock from A3U5 via
A3C82 into the measurement line. It checks on the
amount of 19.2 kHz voltage is developed. If the
developed voltage exceeds a certain level, this is
detected as an open thermocouple with a logic output at
A3U32-7. Similar to the overload detection, the self-test
does not set any channel relays and the detection circuit
should detect the “open” thermocouple. Also check the
relays in the circuit path.
13
Communication
parameters corrupt.
A1 Main
PCA
Background The A1U11 real-time clock device has
resident RAM that stores the RS-232 communication
parameters (baud rate). A1U11 is powered by Vbb, the
source of which is either the battery BT1 or the supply
Vcc, depending if the instrument is powered or not (via
A1U10).
Failure For this error to occur, the RS-232 parameter is
no longer stored in A1U11 RAM or the addressing is
missing.
Correction This might happen if the battery BT1 is dead,
or a problem with A1U10 power supply monitor, or the
A1U11 chip itself. Also check A1U29 for I/O and memory
decoding.
14
Ethernet address
parameter corrupt.
A1 Main
PCA
Background The A1U21 FLASH memory device is
divided into sections. One of these sections is the
memory for the Ethernet address. This is a unique
address assigned at the time of manufacturer.
Failure For this error to occur, the Ethernet address is
either missing or is corrupted and must be reloaded, or
the addressing is missing.
Correction Reloading the Ethernet address is a factory
procedure only. The only recourse is to order a new
A1U21 device programmed at the factory if the Ethernet
address is corrupted. Also see “Troubleshooting the
Digital Kernel.” Also check A1U29 for I/O and memory
decoding.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...