![Fluke NetDAQ 2640A Скачать руководство пользователя страница 168](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217168.webp)
NetDAQ
Service Manual
5-4
•
A good connection with SMT requires only enough solder to make a positive
metallic contact. Too much solder causes bridging, while too little solder can cause
weak or open solder joints. With SMT, the anchoring effect of the through-holes is
missing; solder provides the only means of mechanical fastening. Therefore, the pca
must be especially clean to ensure a strong connection. An oxidized pca pad causes
the solder to wick up the component lead, leaving little solder on the pad itself.
Refer to the Fluke "Surface-Mount Device Soldering Kit" for a complete discussion of
these techniques.
Error Detection
5-3.
At power-up, the instrument software performs self-tests. If any errors in instrument
operation are detected, they are reported on the instrument front panel with Error in the
primary display and a decimal error code number in the secondary display. If there is
more than one error, they are displayed sequentially. Selftest errors can be retrieved from
RS-232 commands and the network.
A selftest includes a test of the following items:
•
FLASH ROM parameters, communication parameters and calibration constants.
•
RAM Instrument and channel configuration plus RAM images of FLASH ROM
parameters.
•
Ethernet Ethernet chip and static Ethernet RAM.
•
Display Display processor and display board
•
Inguard Specific tests for ROM checksum, RAM, A/D converter, zero offset test,
reference balance test, ohms overload test, and otc.
A summary of the possible error codes are shown in Table 5-1, including the front panel
error code (in decimal) and the corresponding network and RS-232 error code (in
hexadecimal). The faults that might cause each error are described in “Troubleshooting
the Instrument” later in this chapter.
Table 5-1. Selftest Error Codes
Front Panel
Error Codes
RS-232 Query
Error Codes
Error Codes in
hexadecimal
Error Code Description
(None)
0
0x00000000
No selftest error.
1
1
0x00000001
Bad boot software image in FLASH ROM.
2
2
0x00000002
Bad main software image in FLASH ROM.
3
4
0x00000004
RAM test failure.
4
8
0x00000008
Display test failure.
5
16
0x00000010
Display not responding.
6
32
0x00000020
Calibration constants corrupt.
7
64
0x00000040
Inguard not responding.
8
128
0x00000080
Inguard A/D failure.
9
256
0x00000100
Inguard zero offsets test failed.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...