![Fluke NetDAQ 2640A Скачать руководство пользователя страница 161](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217161.webp)
Performance Testing and Calibration
Calibration
4
4-39
Table 4-6. Manual VAC Calibration
Command
Response
Action
CAL 2
=>
Puts NetDAQ in VAC calibration.
CAL_REF?
30.0000E-3
5700A - Source 30 mV ac @ 1 kHz
CAL_STEP?
30.0000E-3
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
300.000E-3
5700A - Source 300 mV ac @ 1 kHz
CAL_STEP?
300.000E-3
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
300.000E-3
5700A - Source 300 mV ac @ 1 kHz
CAL_STEP?
300.000E-3
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
3.0
5700A - Source 3V ac @ 1 kHz
CAL_STEP?
3.0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
3.0000E+0
5700A - Source 3V ac @ 1 kHz
CAL_STEP?
3.0000E+0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
30.0000E+0
5700A - Source 30V ac @ 1 kHz
CAL_STEP?
30.0000E+0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
[2640A only for remainder of steps]
CAL_REF?
30.0000E+0
5700A - Source 30V ac @ 1 kHz
CAL_STEP?
30.0000E+0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
300.0000E+0
5700A - Source 300V ac @ 1 kHz
CAL_STEP?
300.0000E+0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
Manual Resistance Calibration Procedure
4-45.
The resistance calibration procedure calculates gain and offset calibration constants for
all of the four-wire ohms ranges. The same calibration constants are used for the
corresponding two-wire ohms ranges.
Complete the following procedure to manually calibrate the ohms function.
1.
If you have not already done so, complete the procedure “Preparing for Calibration”
earlier in this chapter.
2.
Connect the instrument and 5700A Calibrator as shown in Figure 4-8.
3.
Complete the sequence of manual steps shown in Table 4-7. The default values were
chosen assuming that a Fluke 5700A will be used to supply the reference resistors.
For each range, channels 1 and 2 are configured to four-wire ohms and a
measurement is made for each channel. Making a four-wire ohms measurement on
channel 1 provides the full scale reading and measuring the four-wire short on
channels 2 and 12 determines the zero offset error. The four-wire ohms gain and
offset calibration constants are determined from these two measurements.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...