Accessories
Thermostat (A-MP-THERM)
49
7
•
Fix the cooling hoses as shown in the figure above. Ensure that the thermostat (arrow on the housing’s
side) is installed according to the flow direction.
Note
: The fittings needed to connect the inner taper thread of the linescanner tube fitting (Rc 1/8") and
the cooling hose of the Thermostat are not supplied as standard components! We recommend a
® Stainless Reducing Union (8 mm / 0.31 in. outer diameter
– 6 mm / 0.24 in. inner diameter)
•
Move the external temperature sensor into the mounting block and fix it using the M3x8 hex screws Take
care not to damage the sensor when tightening the screws! Do not bend the connection line between
sensitive element and thermostat at a radius tighter than 5 mm (0.2 in)!
•
Adjust the thermostat according to the table below.
Table 7-1: Thermostat Adjustment
Relative Humidity [%]
Am
b
ie
n
t T
e
m
p
e
ra
tu
re
[
°C/
°F
]
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
0/32
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
5/41
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
10/50
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
15/59
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
0
20/68
+
+
+
+
+
+
+
+
+
+
+
+
+
0
0
0
0
0
1
25/77
+
+
+
+
+
+
+
+
+
+
0
0
0
1
1
1
1
1
2
30/86
+
+
+
+
+
+
+
0
0
0
1
1
1
1
2
2
2
2
3
35/95
+
+
+
+
+
0
0
1
1
1
2
2
2
2
3
3
3
3
4
40/104
+
+
+
+
0
1
1
1
2
2
2
3
3
3
4
4
4
4
5
45/113
+
+
0
0
1
2
2
2
3
3
4
4
4
4
5
5
5
5
6
50/122
+
+
0
1
2
2
3
3
4
4
4
5
5
5
6
6
6
6
7
60/140
0
1
2
3
3
4
5
5
5
6
6
7
7
7
7
7
7
7
–
70/158
1
2
4
4
5
6
6
7
7
7
7
7
–
–
–
–
–
–
–
80/176
2
4
5
6
7
7
7
–
–
–
–
–
–
–
–
–
–
–
–
90/194
4
5
7
7
7
–
–
–
–
–
–
–
–
–
–
–
–
–
–
100/212
5
7
7
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
+
thermostat not necessary
0
thermostat recommended
1
– 7
thermostat necessary, 1
– 7: recommended adjustment of the thermostat
–
housing temperature out of specification
•
Check the system functionality. After a few minutes the internal housing temperature should reach the
minimum device temperature as follows:
Thermostat adjustment
Minimum device temperature
0
15°C / 59°F
1
20°C / 68°F
2
25°C / 77°F
3
30°C / 86°F
4
35°C / 95°F
5
40°C / 104°F
6
45°C / 113°F
7
50°C / 122°F
Содержание MP Series
Страница 16: ...MP Series Users Manual Rev 1 0 Apr 2021 16 1024 pixel 80 Hz scan rate Scan Angle FOV All models 90 ...
Страница 18: ...MP Series Users Manual Rev 1 0 Apr 2021 18 2 2 1 2 Far Focus Figure 2 2 Optical Diagrams Far Focus ...
Страница 37: ...Installation Ethernet 37 5 5 Close the dialog box by pressing on Save ...
Страница 44: ...MP Series Users Manual Rev 1 0 Apr 2021 44 7 2 Mounting Plate A MP MP Figure 7 1 Mounting Plate for Tripod ...