Fluke i6000s Flex Series Скачать руководство пользователя страница 1

 

i6000s Flex 

AC Current Probe 

Instruction Sheet 

Introduction 

The i6000s 24 Flex and i6000s 36 Flex AC Current Probes (the 
Probe) are used with oscilloscopes, digital multimeters, recorders 
or data loggers. The Probes can be used to measure AC current 
up to 6000 A. The flexible measuring head allows current 
measurements on conductors that are hard to reach or 
inaccessible using typical clamp-on current probes. 

The Probes provide a low-voltage (3 V ac) output that is 
proportional to the current being measured. The Probes provide 
full-scale direct readings for 60 A, 600 A, and 6000 A. 

Safety Instructions 

A Warning identifies conditions and procedures that are 
dangerous to the user. A Caution identifies conditions and 
procedures that can cause damage to the Product or the 
equipment under test. 



 Warning 

To prevent electric shock or personal injury: 

 

If the Probes are used in a manner not 

specified in these operating instructions, 

protection provided by the Probes may be 

impaired. 

 

Use the Probes only if qualified. 

 

Use caution during installation and use of the 

Probes; high voltages and currents may be 

present in circuit under test. 

 

Have maintenance performed by only qualified 

service personnel. 

 

Protect the Probes against water and 

humidity. 

PN 2842359 

June 2007, Rev. 1, 4/17 

©2007-2017 Fluke Corporation. All rights reserved.  
Specifications are subject to change without notice.  
All product names are trademarks of their respective companies. 

Содержание i6000s Flex Series

Страница 1: ...fies conditions and procedures that are dangerous to the user A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test Warning To prevent electric shock or personal injury If the Probes are used in a manner not specified in these operating instructions protection provided by the Probes may be impaired Use the Probes only if qualified Use cauti...

Страница 2: ...Probe to display device before installing the flexible measuring head Never change batteries while measurement head is installed on conductor Use only the provided original or specified accessories Adhere to local and national safety codes Individual protective equipment must be used to prevent shock and arc blast injury where hazardous live conductors are exposed CAT III equipment is designed to ...

Страница 3: ...ry III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Double Insulated Do not apply around or remove from uninsulated hazardous live conductors without taking additional protective measures Conforms to European Union directives Conforms to relevant Australian Safety and EMC standards Certified by CSA Group to North A...

Страница 4: ...ing Humidity 15 to 85 non condensing Weight 180 g head 190 g electronics Safety Standards General IEC 61010 1 Pollution Degree 2 Measurement IEC 61010 2 032 CAT III 600 V Electromagnetic Compatibility EMC IEC 61326 1 Portable Electromagnetic Environment IEC 61326 2 2 CISPR 11 Group 1 Class A Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequency energy that ...

Страница 5: ...nstruction Sheet Page 5 Operating Instructions ewf0001 eps Measuring head Measuring head coupling Head output cable Enclosure Power on Range selector Battery low indicator Power on indicator Output cable BNC connector ...

Страница 6: ...ed from other equipment To install the battery 1 Use a coin or a similar tool to turn the battery lock turn until the dot aligns with the unlock symbol 2 Remove the battery cover 3 Install the batteries ensuring that correct polarity is observed 4 Replace the battery cover and turn the battery lock until the dot aligns with the lock symbol Measuring Current Warning To prevent electric shock or per...

Страница 7: ...Always use appropriate equipment for personal protection When installed on bare conductors busbars the product must be within a suitable enclosure Ieb0002 eps Operation To turn on the unit move the switch from Off position to the required measurement range If the value of current being measured is unknown select the 6000 A current range and then reduce accordingly Battery Status Battery status is ...

Страница 8: ...sposable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To obtain service during the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a ...

Отзывы: