i400 

AC Current Clamp 

 

 

Calibration Information 

 

Introduction 

 

XW

Warning 

To avoid electric shock or injury, do not perform the performance tests or 
calibration procedures unless you are qualified to do so. 

The information provided in this manual is for the use of qualified personnel 
only. 

 

The 

i400 Calibration Information

 provides the information necessary to verify the performance of the 

Fluke i400 AC Current Clamp, hereafter known as the Current Clamp.  

 

The following information is included in this document: 

 

Safety Information and Electrical Symbols  

 

Specifications  

 

Maintenance 

 

Performance Tests  

 

User-Replaceable Parts 

 

Warranty Statement  

 

See the 

i400 Instruction Sheet 

for complete operating instructions. 

Contact Information 

 

To contact Fluke, call: 

 

USA:  1-888-44-FLUKE (1-888-443-5853) 

Canada:  1-800-36-FLUKE (1-800-363-5853) 

Europe:  +31 402-675-200 

Japan:  +81-3-3434-0181 

Singapore:  +65-738-5655 

Anywhere in the world:  +1-425-446-5500 

 

USA Service: 1-888-99-FLUKE (1-888-993-5853) 

 

For additional information about Fluke, its products, and services, visit Fluke’s web site at:

 

www.fluke.com

 

 

To register this product, go to 

register.fluke.com

 

Содержание i400

Страница 1: ...lamp hereafter known as the Current Clamp The following information is included in this document Safety Information and Electrical Symbols Specifications Maintenance Performance Tests User Replaceable...

Страница 2: ...ages higher than 1000 V CAT III or 600 V CAT IV CAT III equipment is designed to protect against transients in equipment in fixed equipment installations such as distribution panels feeders and short...

Страница 3: ...cifications Reference Conditions 23 5 C 20 to 75 RH 73 41 F 20 to 75 RH conductor centered in jaw opening no dc component no adjacent conductor Measurement Range 1 A to 400 A ac rms Output 1 mA A Accu...

Страница 4: ...fore each use inspect the Current Clamp Look for cracks or missing portions of the clamp housing and output cable insulating cover Look for loose or weakened components Pay particular attention to the...

Страница 5: ...verify the complete operation of the Current Clamp and checks the accuracy of each function against the Clamp s specifications If the Current Clamp fails any part of the test adjustment or repair is i...

Страница 6: ...t 5520A LCOMP Setting UUT Nominal Output Lower Upper 1 1 A 45 Hz 1 A 45 Hz OFF 1 mA 45 Hz 0 920 mA 1 080 mA 2 100 A 400 Hz 2 0 A 400 Hz ON 100 mA 400 Hz 97 94 mA 102 06 mA 3 400 A 60 Hz 8 0 A 60 Hz ON...

Страница 7: ...st Steps 2 and 3 User Replaceable Parts The Current Clamp contains no user replaceable parts Table 4 lists user documentation that is available for this product Table 4 User Documentation Description...

Страница 8: ...authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fl...

Отзывы: