Fluke i200 Скачать руководство пользователя страница 1

®

 

i200/i200s

 

AC Current Clamp 

Instruction Sheet 

Introducing the i200/i200s 

The i200 is a single range 200A clamp-on AC Current Clamp with 
current output via safety shrouded banana plugs.  
The i200s is a dual range 20A and 200A clamp-on AC Current 
Clamp with voltage output via a safety insulated BNC connector.  
A dual banana to BNC adapter is supplied to allow the i200s to be 
connected to multimeters with banana input. 
 

Unpacking 

The following items should be included in your Current Clamp box: 

 Current 

Clamp 

 

Dual Banana to BNC Adapter model PM9081 (only with i200s) 

 

Instruction Sheet (this paper) 

Check the contents of the shipping box for completeness. If 
something in the box has been damaged or missing, contact your 
distributor or the nearest FLUKE sales or service office immediately. 
 

Safety Information 

 

 

Read First: Safety Information. 

To ensure safe operation and service of the 
current clamp, follow these instructions: 

  Read the operating instructions before use and follow all safety 

instructions. 

  Use the Current Clamp only as specified in the operating 

instructions, otherwise the clamp’s safety features may not 
protect you. 

  Adhere to local and national safety codes. Individual protective 

equipment must be used to prevent shock and arc blast injury 
where hazardous live conductors are exposed. 

  Do not hold the Current Clamp anywhere beyond the tactile 

barrier, see Figure 1. 

  Before each use, inspect the Current Clamp. Look for cracks or 

missing portions of the clamp housing or output cable 
insulation. Also look for loose or weakened components. Pay 
particular attention to the insulation surrounding the jaws. 

  Check the magnetic mating surfaces of the clamp jaws; these 

should be free of dust, dirt, rust and other foreign matter. 

  Never use the clamp on a circuit with voltages higher than 600 

V CAT III. 

  CAT III equipment is designed to protect against transients 

in equipment in fixed equipment installations, such as 
distribution panels, feeders and short branch circuits, and 
lighting systems in large buildings. 

  Use extreme caution when working around bare conductors or 

bus bars. Contact with the conductor could result in electric 
shock. 

  Use caution when working with voltages above 60 V dc, 30 V ac 

rms or 42 V ac peak. Such voltages pose a shock hazard. 

 
 
 

 

            

Figure 1. Safely holding the Current Clamp 

 

Symbols 

 

May be used on HAZARDOUS LIVE 
conductors. 

 

Product is protected by double insulation. 

 

 

Risk of Danger. Important information. See 
Instruction Sheet. 

 

Risk of Electric Shock. 

 

Conforms to relevant European standards. 

 

Earth ground. 

 

Specifications 

SAFETY 

 

 

Input jaws & output floating 
voltage to ground 

Complies with American 
industry standards 
UL61010B-1 & UL61010B-2-
032 and European standards 
EN/IEC 61010-1 2nd Edition 
& EN/IEC 61010-02-032 for 
600V CAT III, pollution 
degree 2. 

 

 

 

EMC 

Complies with standards 
EN/IEC 50081-1 &  
EN/IEC 50082-2 

 

 

ELECTRICAL SPECIFICATIONS 

All Electrical Specifications are valid at the following reference 
conditions: 

 

 

Ambient temperature 

23±3°C (73±3°F) 

 

Relative Humidity 

20 to 75% 

 

Frequency 

48 to 65 Hz 

 

Continuous external field 

< 40 A/m 

 

Load impedance 

i200: 0.2

 …15

 

i200s: >1 M

 // 100 pF 

 

The current may not contain any DC component 

 

No influence from adjacent currents 

 

The conductor must be centered within the jaw aperture 

 

20A Range (i200s only) 

Measuring range 

0.1 to 24A 

Maximum current 

24A 

Crest factor * 

< 3 

Maximum non-destructive 
current 

200A (Frequency 

 1 kHz and 

crest factor < 3) 

Output signal 

100 mV/A 

Output impedance 

 20 

 @ 1 kHz 

Basic accuracy 

 

48 Hz to 65 Hz 

 2% + 0.5A 

Additional error: 

40 Hz to 48 Hz and  
65 Hz to 1 kHz 

+ < 10% 

1 kHz to 10 kHz 

+ < 15% 

Phase shift 

Unspecified 

 
200A Range 

i200 

i200s 

Measuring range 

0.5 to 240A 

0.5 to 240A 

Maximum current 

240A 

240A 

Crest factor * 

< 3 

< 3 

Maximum non-destructive 

current 

@ Frequency 

 1 kHz and  

crest factor < 3 

Continuous 200A 
10 min ON 
/30 min OFF 

240A 

Output signal 

1 mA/A 

10 mV/A 

Output impedance 

 10 

 @ 1 kHz 

Basic accuracy 

 

 

48 Hz to 65 Hz 

 

 

0.5A to 10A 

 3% + 0.5A 

 3.5% + 0.5A 

10A to 40A 

 2.5% + 0.5A 

 3% + 0.5A 

40A to 100A 

 2% + 0.5A 

 2.5% + 0.5A 

100A to 240A 

 1% + 0.5A 

 1.5% + 0.5A 

Additional error: 

 

 

40 Hz to 48 Hz and  
65 Hz to 1 kHz 

 
+ < 3% 

 
+ < 3% 

1 kHz to 10 kHz 

+ < 12% 

+ < 12% 

Phase shift 

 

 

0.5A to 10A 

Unspecified 

Unspecified 

10A to 40A 

 5 ° 

 6 ° 

40A to 100A 

 3 ° 

 4 ° 

100A to 240A 

 2.5 ° 

 3 ° 

 

 

All ranges 

i200 

i200s 

Load on output 

0.2...15 

 >1 

M

 // < 100 pF

Load Influence 

Current: < 1%  
Phase: < 1

°

 


Bandwidth -1.5 

dB 

 -3dB 

40 Hz to 10 kHz 
40 kHz 

40 Hz to 10 kHz 
40 kHz 

Additional errors 

 

 

With temperature 

 0.15 % / 10 K 

With position of 
conductor in the 
clamp aperture 

 0.5 % @ 50 Hz 

With adjacent 
conductors 

 15 mA / A @ 50 Hz 

 

 

 

*

  This is the maximum permissible ratio between the peak value 

of a superimposed transient and the ac rms value. 
 

 

GENERAL  

Clamp Dimensions 

135 x 50 x 30 mm  
(5.3 x 2 x 1.2 in) 

Protection index 

IP40 

Jaw Opening 

21 mm (0.82 in) 

Height of open Jaws 

69 mm (2.7 in) 

Maximum conductor size 

 20 mm (0.8 in) or  

busbar 20 x 5 mm (0.8 x 0.2 in) 

Weight 

180 g  (6.4 oz) 

Cable length 

i200 

 i200s 

1.5 m (59 in) 
2m (79 in) 

 

 

Temperature 

Operating 
Non-operating 

 
-10 to +55°C (+14 to +131°F) 
-40 to + 70°C (-40 to +158°F) 

Relative Humidity 

Operating 
 

 
85%, up to +30°C (+86°F) 
75%, up to +55°C (+131°F) 

Altitude 

Operating 
Non-operating 

 
to 2000 m (6500 ft) 
to 12000 m (40000 ft) 

EMC EN/IEC 

50081-1 

EN/IEC 50082-2  (3V/m, 
2.74V/yd) 

 

 

Instrument Compatibility 

The i200s is compatible with any Fluke ScopeMeter test tool, Power 
Harmonics Analyzer, Oscilloscope, Multimeter, or other voltage 
measurement device that has the following features: 

 

BNC input connector. The Dual Banana to BNC Adapter 
included in the package, can be used to connect to standard 
inputs on multimeters. For the 120 series ScopeMeters, use 
the BB120 Shielded Banana to BNC Adapter. 

 

Input accuracy of 2% or better to take full advantage of the 
accuracy of the Current Clamp. 

 

Input impedance of greater than or equal to 1 M

, and for full 

bandwidth and accuracy, a maximum input capacity of 100 pF. 

 

A pass- band of more than four times the frequency of the 
waveform to be measured. 

 
The i200 is compatible with any Fluke Multimeter or any other 
current measurement device that has the following features: 

 Banana 

inputs. 

 

Input accuracy of 2% or better to take full advantage of the 
accuracy of the Current Clamp. 

 

Input impedance of 0.2

 …15

 

 

A pass-band of more than four times the frequency of the 
waveform to be measured. 

 

4822 872 00934 
July 1998, Rev. 5, 08/05 

© 1998 - 2005 Fluke Corporation. 
All rights reserved.  
Printed in France  
All product names are trademarks of 
their respective companies. 

Tactile barrier

 